9-31
Meter Function Menus
9
* The
‘MIN’
button allows you to sequence through BUS.1, BUS.2, BUS.3,
BUS.4, BUS.5, BUS.6, BUS.7 and BUS.8.
The
‘MAX’
button allows you to select the “0” or “1” state for each “BUS”
condition.
The
‘MENU’
button stores the selected values for each “BUS” condition
changed and advances the meter to the next configuration (“SERCNT”).
Every underlined “0” or “1” state is the factory preset value.
9.2.23 SERCNT (Serial Count)
Serial count is used to program the number of readings the meter must take
(programmable from “00001” to “59,999”) between transmissions of data via
the optional RS-232 or RS-485 serial communications board.
Once you are done with the changes, press the
‘MENU’
button to display the
normal operating display. The meter advances to run mode showing the
currently measured values. Pressing the
‘RESET’
button two times allows
you to return to the run mode.
MENU
MIN/MAX/MENU *
BUTTON
BUTTON
MAIN MENU
SUB MENU
DESCRIPTION
“SERCNT”:
SERIAL COUNT
This specifies the number
of readings between data
transmissions:
“00000”
“00001” to “59999”
(“00001”)
* Press the
‘MIN’
button to show the value on the display.
The
‘MIN’
button also allows you to change the position of the digit being
changed.
The
‘MAX’
button changes the value of the digit to be displayed.
Press the
‘MENU’
button to store the changes and advances the meter to
“RESET 2” and returns to the run mode.
The underlined item is the factory preset value.
NOTE:
After you complete your programming you may want to go back and
remove the jumper in S3A position on the main board to insure against any
unauthorized changes.