12
Analog Output Options
12-4
If you are installing an analog output board, follow this “one time only”
procedure:
1.
Write down the four (4) numbers inscribed on the top edge of your
analog output board below:
CAL VZ =
CAL VS =
CALmAZ =
CALmAS =
2.
“Reveal the main board” and install the analog output board using the
procedures outlined in Section 5.2.
3.
Attach connector wires, insert connectors, and apply power to the meter
as described in Section 5.3.
4.
If not already unlocked, press the
‘MENU’
button until “L4 CNF” is
displayed and press the
‘MIN’
button six times. Now press the
‘MAX’
button to set “L4C.6=0”.
5.
Press the
‘MENU’
button to store and advance to “CAL VZ”.
Use the
‘MIN’
and
‘MAX’
buttons to enter the value (recorded from the
edge of the board).
6.
Press the
‘MENU’
button to store and advance to “CAL VS”.
Use the
‘MIN’
and
‘MAX’
buttons to enter the value.
7.
Repeat for “CALmAZ”.
8.
Repeat for “CALmAS”.
9.
Press the
‘MENU’
button to store your entries and then you will see
“C.JUN.OF”. Press the
‘RESET’
button two times and you will see
“RESET2”, followed by “RUN”. Verify your calibration points for the
analog output.