8-14
STRAIN
0
8
Scaling to
Display
Engineering
Units
MENU
SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF
Skip past
L2CNF
Skip past
L3CNF
Skip past
L4CNF
Skip past
INPUT
Skip past
RDG.CNF
Skip past
RDG SC
0.00100
Enter new scale
value from above.
RDG OF
Skip past
(Leave as 000000)
IN CNF
INP.6=1
Activates “IN.SC.OF”
(2-point scaling)
IN.SC.OF
INPUT 1
–0055.6
This is where you
READ 1
00000.0
enter the data
INPUT2
16136.6
obtained during the
READ 2
04000.0
calibration run
(see above).
DEC PT
FF.FFFF
Correct decimal
point selection
CNT BY
Press the ‘RESET’
button two times to
return to the run mode
The unit is now completely scaled. If during verifying the
calibration the display is not correct you can use the offset function
(“RDG OF” if all values read high or low) or perform this calibra-
tion again.