PV8355F4E007 23
5. Notes on Storing
5.1. Storing
See
“
5.1 Storage
”
in the instruction manual for the NIR Micro-spectrophotometer
USPM-W-B.
5.2. Maintenance
See
“
5.2 Maintenance
”
in the instruction manual for the NIR
Micro-spectrophotometer USPM-W-B.
5.3. Wavelength offset
The method used to offset the spectral transmission factor is not provided. Our company
does not perform offset.
For details on how to offset the wavelength, see
“
5.3 Wavelength offset
”
in the
instruction manual for the NIR Micro-spectrophotometer USPM-W-B
.
6. If Abnormality Occurs
Make confirmation through the combined use of
“
6. If Abnormality Occurs
”
in the
instruction manual for NIR Micro-spectrophotometer USPM-W-B.
7. Conforming to Laws and Regulations
See
“
7. Conforming to Laws and Regulations
”
in the instruction manual for the NIR
Micro-spectrophotometer USPM-W-B.
Trouble
Cause and countermeasure
Light is not
output.
(1) Confirm that the lamp has not deteriorated.
(2) Confirm that the lamp is installed securely.
(3) Confirm that the lamp switch is set to ON.
(4) Confirm that the shift stage is adjusted to point 0 on the scale.
(5) Confirm that the position of the (optional) optical path switching
knob is properly set.
(6) Confirm whether the transmission window is dirty.
(See
“
Appendix: Cleaning the transmission window glass.
”
)