PV8355F4E007 24
Appendix
To customers who execute micro-spectrophotometry by placing a sample on the tilt stage
It is recommended to adjust the tilt of the tilt stage vertically to the optical axis when the
customer who purchased a transmit unit and 45-degree expansion unit performs
micro-spectrophotometry. For the customer who measures a thin plate, especially, be sure
to perform this adjustment.
[Tilt adjustment of the tilt stage]
(1) See
“
3.7 Device connection
”
in the instruction manual for the NIR
Micro-spectrophotometer USPM-W-B
for installation.
(2) See
“
4.2 Start-up
”
in the instruction manual for the NIR
Micro-spectrophotometer USPM-W-B
for system start.
(3) Set a revolver.
The revolver is adjusted to the through position (in which an objective lens is not
installed).
(4) Install a BK7 reference sample in the tilt stage.
(5) Adjust the tilt of two axes to the optical axis using the tilt stage adjustment knob.
[Tilt adjustment]
Make adjustment using the tilt adjustment knobs so that outgoing beams are parallel to the
optical axis while confirming a FOCUS window. Adjust such so that the (X:*** Y:***) value
displayed in the lower-left position of the FOCUS window is the minimum. It is
recommended that you set the value to (X:0±1 Y:0±1) when performing more accurate
measurement.
With the tilt out of position
With the tilt adjustment completed
(6) After the tilt adjustment of the tilt stage is completed, measure according to the items
in
“
4. Usage
”
in the instruction manual for the NIR Micro-spectrophotometer
USPM-W-B.
Tilt stage adjuster
Adjust the tilt of the two axes to
the optical axis.