122
Chapter 5 Operation
VIDEO SYSTEM OTV-SI
5.14 Default settings
Users can restore this instrument’s default settings (settings made at the factory
prior to shipment) by loading NO.[ 0 ] as described in Section 5.13, “Load user
settings”. The default settings are given in Table 5.3 below.
Default settings on this instrument
Menu item
Function
Default setting
BRGHT
Exposure level
Center (0)
ENH. [ ]
Image enhancement
1 (LOW)
C.BAR
Color bar
OFF
ASGN.
Remote control switches
SW1: menu execution (EXEC)
SW2: menu selection (MOVE
)
SW3: NONE
SW4: menu selection (MOVE
)
HUE
Color tone level
HUE: center (0)
CHROM
CHROM: center (0)
RED
RED: center (0)
BLUE
BLUE: center (0)
ZOM [ ]
Zoom
Normal (1
) OFF
F [ ]
Freeze mode
FIELD
SHOW
Patient data display
Display (When the videoscope is
connected.)
Not-display (When the camera head is
connected.)
LC [ ]
Brightness control
AUTO
ROT [ ]
Rotated image
Normal
MES [ ]
Iris
FULL
EXIT
Display menu
Displayed (When the videoscope and
OTV-S7H-N/1N/1D are connected.)
Not-displayed (When
OTV-S7H-1D-F08E/L08E/D-L08E are
connected.)
PMP [ ]
Air feed pump
OFF
Table 5.3