![background image](http://html2.mh-extra.com/html/olympus/epoch-650/epoch-650_user-manual_3929357254.webp)
DMTA-10055-01EN, Rev. A, February 2015
Chapter 11
242
Figure 11-14 DGS/AVG setup page
The
DGS/AVG
setup page allows you to define the exact probe being used for the
inspection, and to set up the DGS/AVG curve to be drawn. There are several
adjustments that can be made from this setup page:
—
DGS/AVG
Activate/Deactivate the DGS/AVG function
—
Probe Type
Selects the type of probe to be used (Straight Beam, Angle Beam, or Dual
Probe).
—
Probe Name
Selects the actual probe to be used.
—
Reflector Type
Defines the type of reflector to be used to acquire a reference indication to
build the DGS/AVG curve. For straight beam and dual probes, the available
reflectors are:
•
Back wall
•
Side-drilled hole (SDH)
•
Flat-bottom hole (FBH)
Содержание EPOCH 650
Страница 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Страница 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Страница 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Страница 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Страница 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Страница 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Страница 153: ...DMTA 10055 01EN Rev A February 2015 Programmable Inputs and Outputs 141 Figure 8 1 The A Out setup page...
Страница 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Страница 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Страница 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Страница 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Страница 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Страница 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...