DMTA-10055-01EN, Rev. A, February 2015
Chapter 11
266
11.6.1
Waveform Averaging Setup
If the Waveform Averaging software option has been licensed on the instrument, it
can be activated at any time during standard operation.
To activate the Waveform Averaging software feature
1.
Select
Meas Setup > Special
(group 3/5).
The
Special
screen appears (see Figure 11-28 on page 266).
2.
In the
Special
menu, select
Average
.
3.
Select the level of averaging.
4.
Press
to return to the live screen.
Figure 11-28 Waveform Averaging option setup
11.6.2
Using Waveform Averaging
When Waveform Averaging is on, the
xAv
flag (
) is displayed in the flag area to
alert you that the option is active (see Figure 11-29 on page 267).
Содержание EPOCH 650
Страница 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Страница 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Страница 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Страница 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Страница 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Страница 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Страница 153: ...DMTA 10055 01EN Rev A February 2015 Programmable Inputs and Outputs 141 Figure 8 1 The A Out setup page...
Страница 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Страница 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Страница 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Страница 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Страница 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Страница 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...