13. Once a steady reading is achieved, choose
Auto Cal > Cal-Vel
.
The screen freezes and the
Cal-Vel
box appears (see Figure 9-15 on page 164.
Figure 9-15 Entering the Velocity Cal thickness value
14. Adjust the value to match the known thickness of the gated indication (0.300 in. in
this example), and then choose
Done
to complete the calibration process.
If for any reason you need to exit without acquiring calibration data, press
Cancel
.