DMTA-10055-01EN, Rev. A, February 2015
Managing Special Waveform Functions
115
6. Managing Special Waveform Functions
This chapter describes how to manage special waveform functions of the EPOCH 650
ultrasonic flaw detector.
6.1
Reject
The
Rcvr > Reject
parameter allows you to eliminate unwanted, low-level signals
from the display. The reject function is linear and adjustable from 0 % to 80 % FSH.
Increasing the reject level does not affect the amplitude of the signals above the reject
level.
The reject function can also be used in the unrectified
Rcvr > Rect =
RF
mode.
The reject level is displayed as an horizontal line on the instrument display (see
Figure 6-1 on page 116) or two lines in the case of the unrectified
Rcvr > Rect =
RF
mode.
Содержание EPOCH 650
Страница 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Страница 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Страница 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Страница 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Страница 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Страница 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Страница 153: ...DMTA 10055 01EN Rev A February 2015 Programmable Inputs and Outputs 141 Figure 8 1 The A Out setup page...
Страница 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Страница 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Страница 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Страница 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Страница 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Страница 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...