DMTA-10055-01EN, Rev. A, February 2015
Software Overview
67
3.11.2
Angle Beam Calibration
Angle beam calibrations typically require four steps.
To perform an angle beam calibration
1.
Locate the Beam Index Point (see “Locating the Beam Index Point” on page 165).
2.
Verify the refracted angle of wedge (see “Verifying the Refracted Angle” on
page 167).
3.
Complete the distance calibration using “Measurement calibration” on page 62
(Calibration mode set to
SoundPath
or
Depth
depending on the reflector type).
4.
Set the sensitivity using “Calibrating for Sensitivity” on page 173 (See Figure 3-18
on page 67).
— Capture the sensitivity hole or notch in gate 1, bringing the indication to 80 %
screen height using 2
ND
F, (AUTO XX%).
— Set the reference gain by pressing 2
ND
F, (REF dB).
Figure 3-18 Reference gain setting
Содержание EPOCH 650
Страница 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Страница 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Страница 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Страница 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Страница 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Страница 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Страница 153: ...DMTA 10055 01EN Rev A February 2015 Programmable Inputs and Outputs 141 Figure 8 1 The A Out setup page...
Страница 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Страница 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Страница 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Страница 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Страница 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Страница 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...