6-10
OLYMPIC BILI-LITE
Model 33/34 Instruction Manual
Repair Procedures
7
Reattach the light hood (page 6-5).
8
Install the bulbs (for placement, see page 5-2).
9
Slide the safety shield back into the light hood, then secure it in place with the
retaining screws and washers.
10
Perform the test procedure (see page 2-8).
CAUTION
The safety shield must always be in place during phototherapy to protect the patient from
ultraviolet radiation and broken bulbs.