OLYMPIC BILI-LITE
Model 33/34 Instruction Manual
Replacing the Bulbs
Replacing the Bulbs
Replace the Bili-Lite bulbs at 3,000 hours to maintain greater than 75% of initial
irradiance (irradiance measured at 100 hours).
Bulb Timer
The bulb timer operates whenever the bulbs illuminate. View the elapsed time
shown to track bulb life and determine when to replace the treatment bulbs.
Required Items:
Replacement bulbs (see
Phillips-head screwdriver
To replace the bulbs:
1
Press the power switch to the
OFF
(O) position, then unplug the power cord.
2
Remove the two retaining screws and washers on one end of the safety shield,
and then slide the safety shield out of the light hood.
3
Rotate each individual bulb out of its socket.
4
Install the bulbs: For each new bulb, insert the prongs into the socket, then
carefully rotate the bulb into place. Place the white bulbs on the outside and the
blue bulbs in the center.
5
Slide the safety shield back into the light hood, then secure it in place with the
retaining screws and washers.
6
Test for proper operation (see page 2-8).
NOTE
#
The
elapsed time
bulb timer cannot be reset by the user.
The bulb timer may be labeled as
Lamp Timer
.
Change all treatment bulbs at the same time.
Record the bulb change in the
Medical Device History
.
CAUTION
Use only bulbs specified by Olympic Medical. Use of other bulbs may affect the safety and
efficacy of treatment.