Repair Procedures
b
Starting at one corner, carefully pull the reflector away from the hood. Pull
the sides of the hood outward slightly to allow the reflector to clear the
shield tab.
c
Once the reflector is clear of the hood, unplug wires 5 and 22 from the
power switch, then unclip wires 5 and 22 from the clip centered on the
inside the hood (see
Figure 6.4). It may be possible to lay the reflector down
next to the hood to work on the bulb holder. (In older models, cut the wires
or otherwise detach them.)
Figure 6.4
Inside of hood with reflector removed
To re-attach the wiring:
1
Place wires 5 and 22 into the clip centered on the inside the hood.
2
Insert the reflector into the hood.
3
Place the washers, then tighten the nuts that secure the reflector to the
light hood.
4
Attach the fan shroud: Tighten the three Phillips-head screws and washers.
5
Attach the safety shield bracket: Tighten the two acorn nuts and two
Phillips-head screws and washers.
6
Reattach the light hood (page 6-5).
7
Install the bulbs (for placement, see page 5-2).
8
Slide the safety shield back into the light hood, then secure it in place with the
retaining screws and washers.
9
Perform the test procedure (see page 2-8).
Connector
Clip
CAUTION
The safety shield must always be in place during phototherapy to protect the patient from
ultraviolet radiation and broken bulbs.