711449 (Rev.3) 02/2016
MiniOX
®
3000 Oxygen Monitor
Section 2, Setting up the MiniOX
®
3000 Oxygen Monitor
Section 2
Setting Up the MiniOX
®
3000
Oxygen Monitor
The MiniOX
®
3000 Oxygen Monitor is packaged in
a shipping carton suitable for instrument storage. If
there is obvious damage to the shipping carton or its
contents, contact Ohio Medical Corporation Repair at:
1-847-855-0800
To set up the monitor:
1. Verify that you have the following (see Figure
2-1):
• MiniOX
®
3000 Oxygen Monitor
•
2[\JHQVHQVRUDQGGHÀHFWRULQVHDOHG
package
• 10-foot coiled cable with twist collar.
• Tee adapter
• Sensor retaining strap
• 9-volt alkaline battery
• MiniOX
®
3000 Operation Manual CD
2. Check the sensor’s manufacturing date on the
sensor package. If the sensor is put into service
within six months after this date, the sensor will
PHHWSHUIRUPDQFHVSHFL¿FDWLRQV
3. Remove the sensor from the sealed package
and attach it to the coiled cable.
a. Firmly press the connector until it snaps
into place; tighten the twist collar.
b. Insert the opposite end of the coiled cable
into the jack on the side panel of the
instrument; tighten the twist collar.
4.
5HPRYHWKHGHÀHFWRUIURPWKHSDFNDJH
a. Insert the gasket into the open end of
WKHGHÀHFWRUHQVXULQJWKDWWKHJDVNHWLV
SURSHUO\VHDWHGZLWKLQWKHGHÀHFWRU
Figure 2-1
MiniOX
®
3000 Oxygen Monitor
OXYGEN SENSOR AND DEFLECTOR
IN SEALED CONTAINER
9-VOLT
ALKALINE
BATTERY
SENSOR
RETAINING STRAP
TEE
ADAPTER
MINIOX
®
3000
OXYGEN MONITOR
COILED
CABLE
TWIST
COLLAR
2-1