background image

711449 (Rev.3)  02/2016

NOTE:  To maximize battery life, press l/0 to 

turn OFF the MiniOX

®

 3000 unit when 

not monitoring. In order to retain alarm 
settings, do not remove battery.

Sensor Replacement 

To ensure safe and effective use of your device, the 
sensor must be replaced with a MiniOX

®

 sensor as 

this sensor is manufactured for this instrument. Use 
of other types of sensors has not been tested and 
is not endorsed by Ohio Medical Corporation. Use 
of other sensor types will void your warranty.

There are no serviceable parts in the sensor or 
cable assemblies, the entire unit must be replaced.

Replace the sensor when: 

•  Room air reading is greater than 20.8% ±2% 

(18.8% to 22.8%) in Two-Point Linearity Check. 

•  The MiniOX

®

 3000 Oxygen Monitor will not 

calibrate. 

•         and “OFF” display and audible and 

visual alarms persist when sensor and cable 
connections are correct and cable is viable. 

To replace the sensor: 

1.  Verify that the monitor is turned OFF. The display 

should be blank. 

2.  Disconnect the expired sensor from the coiled 

cable. 

3. 

$WWDFKDQHZVHQVRUWRWKHFRLOHGFDEOHDQG¿UPO\

press the connector until the sensor snaps into 
place. Tighten the twist collar. 

4.  Recalibrate the monitor. 

'HÀHFWRU5HSODFHPHQW

7RUHSODFHWKHGHÀHFWRU

1. 

*HQWO\XQVFUHZWKHROGGHÀHFWRUDQGUHPRYHWKH

gasket. 

2.  Insert a new gasket into the large end of the new 

GHÀHFWRU&DUHIXOO\VFUHZWKHQHZGHÀHFWRURQWR

the sensor, making sure that the new gasket is 
properly seated. 

6HH)LJXUHIRUSURSHUVHQVRUDQGGHÀHFWRU

assembly. 

Cable Replacement 

To replace the coiled cable: 

1.  Verify that the monitor is turned OFF. The display 

should be blank. 

2.  Disconnect the old cable from the sensor. 

3.  Disconnect the old cable from the instrument. 

4.  Attach the new cable to the sensor. Firmly press 

the connector until the sensor snaps into place. 
Tighten the twist collar. 

5.  Attach the new cable to the instrument. Firmly 

press the connector until it snaps into place. 
Tighten the twist collar. 

6.  Recalibrate the instrument. 

Cleaning, Disinfection and Sterilization

Never autoclave, immerse, or expose the 
MiniOX

®

 3000 Oxygen Monitor (including 

sensor) to high temperatures (>70

o

C). Never 

expose the device to pressure, irradiation, 
vacuum, steam, or chemicals (other than 
alcohol or mild cleaning agents). 

Clean the instrument and sensor by wiping with a 
cloth lightly dampened with Isopropyl Alcohol or mild 
detergent. Make sure that no moisture seeps into the 
instrument case or cable jack port. 

Instrument 

When cleaning or disinfecting the instrument, care 
must be taken to prevent entry of solutions into the 
instrument case. If it is suspected that solutions 
or moisture enters the case, verify performance 
by conducting a self-test. (see

 Appendix D, 

Troubleshooting

)

Cleaning 

The external surfaces of the unit may be cleaned 
by wiping them with a cloth moistened with a mild 
detergent solution. 

Disinfection 

The external surfaces of the unit may be disinfected 
by wiping them with a cloth moistened with ethanol 
or Cidex. The instrument is not designed to withstand 
the conditions imposed by steam, ethylene oxide or 
radiation sterilization. 

MiniOX

®

 3000 Oxygen Monitor   

 

 

 

 

               Section 4, Maintenance and Care

CAUTION

4-2

Содержание MiniOX 3000

Страница 1: ...T l fax 1 608 222 6028 Manuale d istruzioni Assistenza al cliente 1 847 855 0800 Manutenzione 1 847 855 0800 Fax 1 608 222 6028 FEDERAL U S LAW RESTRICTS THIS PRODUCT TO SALE BY OR ON THE ORDER OF A P...

Страница 2: ...HQ LGLRP DV 3DUD HQFRQWUDU OD VHFFLRQ HVF ULWD HQ HO LGLRPD GH VX SUHIHUHQFLD SRU IDYRU HOLMD HO VHSDUDGRU FRQ HO FRORU LQGLFDGR D FRQWLQXDFLRQ H PDQXHO FRQWLHQW GHV LQVWUXFWLRQV HQ ODQJXHV 3RXU WURX...

Страница 3: ...o these products are voided if the products are not installed used and serviced in accordance with the instructions in this manual Please protect yourself and your patients by following them We encour...

Страница 4: ...ng warranties shall not apply if the product has been repaired other than by Ohio Medical or in accordance with written instructions provided by Ohio Medical or altered by anyone other than Ohio Medic...

Страница 5: ...the unit OFF and contact your nearest Ohio Medical Service Center for additional information 8 Use of devices generating or emitting electromagnetic radiation near the MiniOX 3000 Oxygen Monitor may i...

Страница 6: ...MiniOX 3000 Oxygen Monitor 2 1 Figure 2 1 MiniOX 3000 Oxygen Monitor 2 1 CAUTION 2 2 Figure 2 2 Assembled MiniOX 3000 Oxygen Monitor 2 2 Section 3 Operation 3 1 Calibration 3 1 CAUTION 3 1 To Calibrat...

Страница 7: ...sor Replacement 4 2 H HFWRU 5HSODFHPHQW 4 2 Cable Replacement 4 2 Cleaning Disinfection and Sterilization 4 2 CAUTION 4 2 Instrument 4 2 Sensor and Cable 4 3 6HQVRU H HFWRU 5HWDLQLQJ 6WUDS DQG 7HH GDS...

Страница 8: ...ases and Vapors C 1 Table C 1 Interferent Gases and Vapors C 1 Appendix D Troubleshooting D 1 Table D 1 Troubleshooting D 1 Appendix E Error Codes E 1 Table E 1 Error Codes E 1 Appendix F Accessories...

Страница 9: ...rm and 18 to 99 low alarm or the default high low settings may be used 50 and 18 respectively Audible and visual alarms activate when oxygen concentrations fall below the preset or default low alarm s...

Страница 10: ...ical current displaying O2 values The Ohio Medical Corporation oxygen sensor is self zeroing when no oxygen is present no current is produced 0 oxygen is displayed The sensor has a minimal response to...

Страница 11: ...using a calibration gas other than room air or 100 oxygen or if the microprocessor detects a calibration error during normal operation WKH GLVSOD DVKHV D ZDUQLQJ PHVVDJH 1 3 each time the monitor is...

Страница 12: ...ons a test function that allows the operator to ensure that the high and low alarms activate at the preset alarm levels See Section 3 Test Functions Radio Frequency Electromagnetic Insensitivity The M...

Страница 13: ...ration Manual CD 2 Check the sensor s manufacturing date on the sensor package If the sensor is put into service within six months after this date the sensor will PHHW SHUIRUPDQFH VSHFL FDWLRQV 3 Remo...

Страница 14: ...sor unnecessarily during calibration or use Body heat can cause the sensor s thermistor to change disproportionately to the change in gas sample temperature at the sensing electrode This can produce s...

Страница 15: ...WUDWLRQ RI When a more precise measurement is desired repeat calibration using 100 oxygen concentration Changes in pressure humidity or temperature may affect calibration accuracy See Appendix B Effec...

Страница 16: ...at 100 O2 Figure 3 2 Note Prior to calibrating at 100 O2 the MiniOX 2 JHQ 0RQLWRU PXVW UVW EH FDOLEUDWHG in room air 1 Calibrate in room air See Section 3 To Calibrate In Room Air 2 Expose the sensor...

Страница 17: ...ibrate in room air 2 Place the sensor in a stream of oxygen with a known concentration of 100 until the reading stabilizes 3 Recalibrate the monitor 4 After calibration measure room air This reading s...

Страница 18: ...eactivates If the alarm condition is corrected and then reoccurs within this interval both audible and visual alarms activate 2 To manually terminate the silence mode press the down DUURZ ODEHOHG ZLWK...

Страница 19: ...Medical Corporation service During calibration CAL ERR DVKHV YLVXDO DQG DXGLEOH DODUPV DFWLYDWH WKHQ DVKHV Calibration performed using a gas other than room air or 100 oxygen or wrong calibration key...

Страница 20: ...m condition Installing the Sensor in a Breathing Circuit To use the sensor in a breathing circuit you will need the following 6HQVRU ZLWK GH HFWRU Tee adapter Retaining strap To install the sensor in...

Страница 21: ...711449 Rev 3 02 2016 SENSOR CABLE RETAINING STRAP SENSOR GASKET DEFLECTOR O RING 15MM TEE Figure 3 3 Installing Sensor into Tee Adapter MiniOX 3000 Oxygen Monitor Section 3 Operation 3 7...

Страница 22: ...re 3 4 Sensor in Breathing Circuit FASTEN RETAINING STRAP TO POSTS ON TEE 15MM TEE ADAPTER DEFLECTOR SENSOR SENSOR CABLE RETAINING STRAP SLIDE DOVETAIL COMPLETELY INTO BRACKET MOUNTING BRACKET INSTALL...

Страница 23: ...battery cover is properly VHDWHG DQG DW RQ WKH EDFN RI WKH 0LQL2 3000 Oxygen Monitor case 7 Recalibrate the monitor Reset the low and high alarms if desired Section 4 Maintenance and Care Use only gen...

Страница 24: ...RU assembly Cable Replacement To replace the coiled cable 1 Verify that the monitor is turned OFF The display should be blank 2 Disconnect the old cable from the sensor 3 Disconnect the old cable from...

Страница 25: ...are used Sterilization 7KH VHQVRU GH HFWRU UHWDLQLQJ VWUDS DQG WHH DGDSWHU may be sterilized using Cidex per manufacturer s instructions steam or ethylene oxide Due to the varying conditions imposed...

Страница 26: ...n sealed package Battery Life Approximately 1500 device hours assuming 80 ON 20 OFF Operating Temperature Range 0o C to 40o C 32o F to 104o F Storage Temperature 40o C to 70o C 40o F to 158o F Humidit...

Страница 27: ...9ROWDJH XFWXDWLRQV LFNHU HPLVVLRQV IEC 61000 3 3 Not applicable Not applicable Guidance and manufacturer s declaration electromagnetic immunity The MiniOX LV LQWHQGHG IRU XVH LQ WKH HOHFWURPDJQHWLF H...

Страница 28: ...ould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency...

Страница 29: ...nce by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the MiniOX 3000 as recommended below according to the maximum output power of the communi...

Страница 30: ...creases the actual concentration of oxygen Humidity in a sample has the same effect as diluting the sample with another gas For example if 100 oxygen is satu rated with 100 humidity the actual concent...

Страница 31: ...NT VOLUME DRY INTERFERENCE EQUIVALENT OF O2 Carbon Dioxide 10 0 1 Cyclopropane 50 0 2 HV XUDQH 7 5 0 1 Diethyl Ether 20 0 3 Q XUDQH 4 0 1 Halothane 5 0 2 Helium 80 0 2 VR XUDQH 3 0 5 0HWKR XUDQH 4 0 1...

Страница 32: ...sensor If alert persists following sensor replacement call Ohio Medical Corporation service During operation CAL DVKHV Sensor signal is out of monitoring range Recalibrate the monitor ensure proper ca...

Страница 33: ...es not reoccur press set 3 The battery is disconnected If an error reoccurs the monitor is inoperative and must not be used call Ohio Medical Corporation Repair at 1 847 855 0800 The error code that a...

Страница 34: ...ration Manual CD 600700 Operating Manual printed 711449 MiniOX 3000 Oxygen Monitor 711451 Battery Cover Assembly 710974 Monitor Stand 655033 H HFWRU 803229 Dovetail 474606 Screw 4x1 4 set of 4 thread...

Отзывы: