711449 (Rev.3) 02/2016
MiniOX
®
2[\JHQ0RQLWRU
$SSHQGL[$6SHFL¿FDWLRQV
A-1
Appendix A
6SHFL¿FDWLRQV
7DEOH$6SHFL¿FDWLRQV
O
2
Alarm
System
LOW/HIGH O
2
ALARM
3XOVDWLQJDODUPÀDVKLQJ/('V
Accuracy
+ 1% of full scale (RTP*) as calibrated in the environment of use
(total accuracy, including linearity, is 2%)
Linearity
+ 1% of full scale (RTP*)
Nominal Response Time (O
2
directed at sensor at 2 liters per
minute at RTP)
90% in 20 seconds
97% in 30 seconds
Sensor Life
Over one year in normal medical conditions
Sensor Shelf Life
Six months minimum (stored in sealed package)
Battery Life
Approximately 1500 device hours, assuming 80% ON, 20% OFF
Operating Temperature Range
0
o
C to 40
o
C (32
o
F to 104
o
F)
Storage Temperature
-40
o
C to 70
o
C (-40
o
F to 158
o
F)
Humidity
5% to 95% RH (non-condensing)
Dimensions
INSTRUMENT
152 x 83 x 33 mm (5.98 x 3.26 x 1.31 in.)
SENSOR
OD 30.56 mm (1.2 in.); Length: 43.2 mm (1.7 in.)
Battery Alarm
System
STAGE 1
Warning that approximately six hours of monitoring life remain; LOW BAT
appears on display; beep sounds every 30 seconds until battery is replaced or
Stage 2 occurs.
STAGE 2
Alarm that proper operation is not assured; monitor shuts down, audible and
visual alarms are activated.
*RTP: Room temperature and pressure, e.g., 23
o
C + 3
o
and ambient barometric
pressure