background image

711449 (Rev.3)  02/2016

MiniOX

®

 3000 Oxygen Monitor   

 

 

 

 

                

      Section 1, Introduction

The operator can silence the audible alarm for 
three 30-second  intervals for a total of 90 seconds; 

KRZHYHUWKHYLVXDODODUPFRQWLQXHVWRÀDVK$W

the end of the silence period, the audible alarm 
reactivates if the alarm conditions is not corrected.  
(

See Section 3, Silencing an Alarm)

Low Battery Alarms

The MiniOX

®

 3000 Oxygen Monitor features a two-

stage battery alarm that warns of depleted and 
expired battery voltage:

• 

7KH¿UVWDODUPDOHUWVWKHRSHUDWRUWKDWWKH

monitor has approximately six hours of 
operating time remaining:

•  a warning message appears in the display

•  an audible alarm sounds at 30-second 

intervals

•  If the operator does not replace the battery 

after this alarm, a second low battery alarm 
activates when the battery is no longer able to 
support monitoring.  The monitor:

•  displays a warning message

•  activates an audible and visual alarm (

See 

Section 3, Low Battery Alarms

)

Sensor Indicator

During monitor, the MiniOX

®

 3000 Oxygen Monitor 

displays a warning message; audible and visual 
alarms activate if:

•  the oxygen sensor becomes disconnected

•  the cable fails or disconnects from the sensor 

or instrument

•  the sensor membrane is perforated

•  the thermistor circuit opens

Calibration Needed Indicator

The MiniOX

®

2[\JHQ0RQLWRUÀDVKHV³&$/´

•  when the monitor is turned ON

•  following sensor disconnection/reconnection.

If a calibration is performed using a calibration 
gas other than room air or 100% oxygen, or if the 
microprocessor detects a calibration error during 
normal operation:

• 

WKHGLVSOD\ÀDVKHVDZDUQLQJPHVVDJH³&$/´

1-3

•  each time the monitor is turned ON 

•  following sensor disconnection/reconnection 

•  when environmental conditions (presssure, 

humidity and temperature) change

When a MiniOX

®

 3000 Oxygen Monitor with a sensor 

connected is turned ON, the monitor performs a self-
test and a calibration requirement occurs. 

• 

³&$/´ÀDVKHVRQWKHGLVSOD\WHOOLQJWKH

operator to carry out the calibration procedure. 

The calibration procedure is initiated by exposing the 

VHQVRUWRDFDOLEUDWLRQJDV>HLWKHUURRPDLUGH¿QHGDV

oxygen concentration of 20.8%) or 100% oxygen] and 
pressing the corresponding key on the keypad.  The 
following appears on the display:

• 

•  the selected calibration gas concentration.

When the      key is pressed the monitor displays:

•  “CAL”

•  a 10-segment bar graph which “counts down” 

two seconds per bar for 20 seconds.

The monitor automatically calibrates to the selected 
concentration.  At the end of the 20 seconds, the 
monitor:

•  emits a beep indicating that calibration is 

complete

•  enters the operating mode

See Section 3, Calibration

Low/High Oxygen Concentration Alarms

The MiniOX

®

 3000 Oxygen Monitor has audible 

and visual alarms that activate when oxygen 
concentrations exceed preset low or high alarm 
settings. Default settings are 18% and 50%, 
respectively. 

When the MiniOX

®

 3000 Oxygen Monitor detects an 

oxygen concentration that exceeds the preset alarm 
limit:

• 

WKHUHG/('IRUWKDWDODUPÀDVKHV

•  an audible alarm activates

•  the measured concentration appears in the 

display

Содержание MiniOX 3000

Страница 1: ...T l fax 1 608 222 6028 Manuale d istruzioni Assistenza al cliente 1 847 855 0800 Manutenzione 1 847 855 0800 Fax 1 608 222 6028 FEDERAL U S LAW RESTRICTS THIS PRODUCT TO SALE BY OR ON THE ORDER OF A P...

Страница 2: ...HQ LGLRP DV 3DUD HQFRQWUDU OD VHFFLRQ HVF ULWD HQ HO LGLRPD GH VX SUHIHUHQFLD SRU IDYRU HOLMD HO VHSDUDGRU FRQ HO FRORU LQGLFDGR D FRQWLQXDFLRQ H PDQXHO FRQWLHQW GHV LQVWUXFWLRQV HQ ODQJXHV 3RXU WURX...

Страница 3: ...o these products are voided if the products are not installed used and serviced in accordance with the instructions in this manual Please protect yourself and your patients by following them We encour...

Страница 4: ...ng warranties shall not apply if the product has been repaired other than by Ohio Medical or in accordance with written instructions provided by Ohio Medical or altered by anyone other than Ohio Medic...

Страница 5: ...the unit OFF and contact your nearest Ohio Medical Service Center for additional information 8 Use of devices generating or emitting electromagnetic radiation near the MiniOX 3000 Oxygen Monitor may i...

Страница 6: ...MiniOX 3000 Oxygen Monitor 2 1 Figure 2 1 MiniOX 3000 Oxygen Monitor 2 1 CAUTION 2 2 Figure 2 2 Assembled MiniOX 3000 Oxygen Monitor 2 2 Section 3 Operation 3 1 Calibration 3 1 CAUTION 3 1 To Calibrat...

Страница 7: ...sor Replacement 4 2 H HFWRU 5HSODFHPHQW 4 2 Cable Replacement 4 2 Cleaning Disinfection and Sterilization 4 2 CAUTION 4 2 Instrument 4 2 Sensor and Cable 4 3 6HQVRU H HFWRU 5HWDLQLQJ 6WUDS DQG 7HH GDS...

Страница 8: ...ases and Vapors C 1 Table C 1 Interferent Gases and Vapors C 1 Appendix D Troubleshooting D 1 Table D 1 Troubleshooting D 1 Appendix E Error Codes E 1 Table E 1 Error Codes E 1 Appendix F Accessories...

Страница 9: ...rm and 18 to 99 low alarm or the default high low settings may be used 50 and 18 respectively Audible and visual alarms activate when oxygen concentrations fall below the preset or default low alarm s...

Страница 10: ...ical current displaying O2 values The Ohio Medical Corporation oxygen sensor is self zeroing when no oxygen is present no current is produced 0 oxygen is displayed The sensor has a minimal response to...

Страница 11: ...using a calibration gas other than room air or 100 oxygen or if the microprocessor detects a calibration error during normal operation WKH GLVSOD DVKHV D ZDUQLQJ PHVVDJH 1 3 each time the monitor is...

Страница 12: ...ons a test function that allows the operator to ensure that the high and low alarms activate at the preset alarm levels See Section 3 Test Functions Radio Frequency Electromagnetic Insensitivity The M...

Страница 13: ...ration Manual CD 2 Check the sensor s manufacturing date on the sensor package If the sensor is put into service within six months after this date the sensor will PHHW SHUIRUPDQFH VSHFL FDWLRQV 3 Remo...

Страница 14: ...sor unnecessarily during calibration or use Body heat can cause the sensor s thermistor to change disproportionately to the change in gas sample temperature at the sensing electrode This can produce s...

Страница 15: ...WUDWLRQ RI When a more precise measurement is desired repeat calibration using 100 oxygen concentration Changes in pressure humidity or temperature may affect calibration accuracy See Appendix B Effec...

Страница 16: ...at 100 O2 Figure 3 2 Note Prior to calibrating at 100 O2 the MiniOX 2 JHQ 0RQLWRU PXVW UVW EH FDOLEUDWHG in room air 1 Calibrate in room air See Section 3 To Calibrate In Room Air 2 Expose the sensor...

Страница 17: ...ibrate in room air 2 Place the sensor in a stream of oxygen with a known concentration of 100 until the reading stabilizes 3 Recalibrate the monitor 4 After calibration measure room air This reading s...

Страница 18: ...eactivates If the alarm condition is corrected and then reoccurs within this interval both audible and visual alarms activate 2 To manually terminate the silence mode press the down DUURZ ODEHOHG ZLWK...

Страница 19: ...Medical Corporation service During calibration CAL ERR DVKHV YLVXDO DQG DXGLEOH DODUPV DFWLYDWH WKHQ DVKHV Calibration performed using a gas other than room air or 100 oxygen or wrong calibration key...

Страница 20: ...m condition Installing the Sensor in a Breathing Circuit To use the sensor in a breathing circuit you will need the following 6HQVRU ZLWK GH HFWRU Tee adapter Retaining strap To install the sensor in...

Страница 21: ...711449 Rev 3 02 2016 SENSOR CABLE RETAINING STRAP SENSOR GASKET DEFLECTOR O RING 15MM TEE Figure 3 3 Installing Sensor into Tee Adapter MiniOX 3000 Oxygen Monitor Section 3 Operation 3 7...

Страница 22: ...re 3 4 Sensor in Breathing Circuit FASTEN RETAINING STRAP TO POSTS ON TEE 15MM TEE ADAPTER DEFLECTOR SENSOR SENSOR CABLE RETAINING STRAP SLIDE DOVETAIL COMPLETELY INTO BRACKET MOUNTING BRACKET INSTALL...

Страница 23: ...battery cover is properly VHDWHG DQG DW RQ WKH EDFN RI WKH 0LQL2 3000 Oxygen Monitor case 7 Recalibrate the monitor Reset the low and high alarms if desired Section 4 Maintenance and Care Use only gen...

Страница 24: ...RU assembly Cable Replacement To replace the coiled cable 1 Verify that the monitor is turned OFF The display should be blank 2 Disconnect the old cable from the sensor 3 Disconnect the old cable from...

Страница 25: ...are used Sterilization 7KH VHQVRU GH HFWRU UHWDLQLQJ VWUDS DQG WHH DGDSWHU may be sterilized using Cidex per manufacturer s instructions steam or ethylene oxide Due to the varying conditions imposed...

Страница 26: ...n sealed package Battery Life Approximately 1500 device hours assuming 80 ON 20 OFF Operating Temperature Range 0o C to 40o C 32o F to 104o F Storage Temperature 40o C to 70o C 40o F to 158o F Humidit...

Страница 27: ...9ROWDJH XFWXDWLRQV LFNHU HPLVVLRQV IEC 61000 3 3 Not applicable Not applicable Guidance and manufacturer s declaration electromagnetic immunity The MiniOX LV LQWHQGHG IRU XVH LQ WKH HOHFWURPDJQHWLF H...

Страница 28: ...ould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency...

Страница 29: ...nce by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the MiniOX 3000 as recommended below according to the maximum output power of the communi...

Страница 30: ...creases the actual concentration of oxygen Humidity in a sample has the same effect as diluting the sample with another gas For example if 100 oxygen is satu rated with 100 humidity the actual concent...

Страница 31: ...NT VOLUME DRY INTERFERENCE EQUIVALENT OF O2 Carbon Dioxide 10 0 1 Cyclopropane 50 0 2 HV XUDQH 7 5 0 1 Diethyl Ether 20 0 3 Q XUDQH 4 0 1 Halothane 5 0 2 Helium 80 0 2 VR XUDQH 3 0 5 0HWKR XUDQH 4 0 1...

Страница 32: ...sensor If alert persists following sensor replacement call Ohio Medical Corporation service During operation CAL DVKHV Sensor signal is out of monitoring range Recalibrate the monitor ensure proper ca...

Страница 33: ...es not reoccur press set 3 The battery is disconnected If an error reoccurs the monitor is inoperative and must not be used call Ohio Medical Corporation Repair at 1 847 855 0800 The error code that a...

Страница 34: ...ration Manual CD 600700 Operating Manual printed 711449 MiniOX 3000 Oxygen Monitor 711451 Battery Cover Assembly 710974 Monitor Stand 655033 H HFWRU 803229 Dovetail 474606 Screw 4x1 4 set of 4 thread...

Отзывы: