711449 (Rev.3) 02/2016
MiniOX
®
3000 Oxygen Monitor
Section 2, Setting up the MiniOX
®
3000 Oxygen Monitor
b.
*HQWO\VFUHZWKHGHÀHFWRURQWRWKHVHQVRU
(For monitoring in a breathing circuit,
see Section 3, Installing the Sensor in a
Breathing Circuit)
5. Unscrew the two battery cover screws on the
back of the instrument.
a. Remove the cover.
b. Install the battery (
See Section 4, Battery
Replacement
for full battery installation
instructions.)
After battery installation:
•
WKHGLVSOD\ÀDVKHV
• the instrument performs a self-test
• then turns OFF
6. Proceed to
Section 3, Operation.
Do not handle the sensor unnecessarily
during calibration or use. Body heat can
cause the sensor’s thermistor to change
disproportionately to the change in gas
sample temperature at the sensing electrode.
This can produce some error until thermal
equilibrium is restored.
See Appendix F, Accessories/Replacement Parts
for
ordering MiniOX
®
3000 Oxygen Monitor accessories
and replacement parts (Figure 2-2).
Figure 2-2
Assembled MiniOX
®
3000 Oxygen Monitor
AUDIBLE ALARMS
LED ALARMS
DISPLAYED OXYGEN
READING
ON/OFF KEY
BATTERY TEST AND
ALARM TEST KEY
21% CALIBRATION KEY
UNLOCK KEY (ENABLES
CALIBRATION SEQUENCE)
DECREASE ALARM SET
POINT KEY
SET KEY (ENABLES
ADJUSTMENTS OF ALARM
SET POINTS)
100% CALIBIRATION KEY
INCREASE ALARM SET
POINT KEY
SILENT KEY
CABLE
ASSEMBLY
MONITOR
STAND
SENSOR
DEFLECTOR
2-2
CAUTION