711449 (Rev.3) 02/2016
Error Codes
The MiniOX
®
3000 Oxygen Monitor performs diagnostic
tests to detect errors that could cause unreliable
monitor operation. These tests are performed:
• when a battery is installed (power-up)
• when the monitor is turned ON
• during routine monitoring
During self-test or any time the monitor detects an
operational error, the MiniOX
®
3000 Oxygen Monitor:
• ceases operation
• display “ERR” and an error code
• activates visual and audible alarms
Error alarms continue until:
1. The monitor is shut off (press
l/0
)
2. The monitor is reset and error does not reoccur
(press set)
3. The battery is disconnected
If an error reoccurs, the monitor is inoperative and must
not be used; call Ohio Medical Corporation Repair at:
1-847-855-0800
The error code that appears on the display corresponds
WRDVSHFL¿FIDLOXUH7KHDXGLEOHDODUPLVDEHHSWKDW
corresponds to this code (e.g., “ERR 3” triggers a
three-beep alarm).
Table E-1. Error Codes
ERROR DISPLAY
CORRESPONDING ERROR
01
System error
02
Random Access Memory (RAM) error
03
Read Only Memory (ROM) error
04
Analog error
05
Timing error
06
Keypad error*
07
Battery error **
08
LCD error
*Press SET. If error reoccurs, call Ohio Medical Corporation Repair
**Replace battery. If error reoccurs, call Ohio Medical Corporation Repair
MiniOX
®
3000 Oxygen Monitor
Appendix E, Error Codes
E-1