APPENDIX B SOFTWARE SERVICE TOOL INSTRUCTIONS
MB90 MB120 Series Service Manual
74
Ohaus Corporation www.ohaus.com
B.3 Replace Control Board.
If replacing a new Control Board to the Analyzer
one will need to perform the 'Replace Control
Board' function.
Note: Kindly take note at the right hand remark in RED with Picture as shown below before
proceeding.
The new Control Board will be pre-loaded with the Drying firmware but to make sure the
firmware is updated one needs to upload the Analyzer firmware via the USB before performing
this function. Updated firmware is available in OH DMX. Kindly refer to Appendix C on how to
install MB90 and MB120 firmware.
Information such as below will need to provide to OHAUS in order to obtain the scale eif file.
1. Analyzer Model Number.
2. Analyzer Serial Number.
3. Analyzer Load Cell Serial Number.