APPENDIX B SOFTWARE SERVICE TOOL INSTRUCTIONS
MB90 MB120 Series Service Manual
79
Ohaus Corporation www.ohaus.com
B.4.2 Linearization (Service Linearization Calibration)
Selecting this function will allow the Analyzer to perform the Service Linearization calibration.
Follow instruction on the Analyzer display screen when selecting this function.
During Service Linearity Calibration below test weight is required.
Model
MB90
MB120
Test weight 25g and 50g
50g and 100g
Note: Do not cover the lid during any weight calibration