APPENDIX C DOWNLOAD ANALYZER FIRMWARE
MB90 MB120 Series Service Manual
83
Ohaus Corporation www.ohaus.com
- Analyzer will skip the firmware upload if its detected the files in the thumb drive is older
version then the firmware uploaded in the Main PCB and Control Board.
5. The whole process will take around 30 sec to complete and during the process the
Analyzer will have a onetime Auto Boot.
6. When you see the below screen only you remove the thumb drive and
RE-BOOT
the
Analyzer to complete the firmware upload.
C2.2 software version SR1.10 or later
Obtain Analyzer firmware from DMX as step 1 in section C2.1
1, Turn on the analyzer
2, Plug the *
USB
to the back of the analyzer 'USB Host com port'
3, Press the 'Setup button' to access the Setup menu, select 'System and data management' for
the 'software update'