CHAPTER 3 – MAINTENANCE PROCEDURES
MB90 MB120 Series Service Manual
46
Ohaus Corporation www.ohaus.com
5. Use both fore fingers to spread the clamping spring to the left and right and at the
same time use both thumbs to lift the Halogen lamp carefully upward out of the
clamping springs.
6. Remove and replace a defective halogen lamp.
Note:
The two ends of the Halogen lamp must not be bent toward or away from each other;
otherwise, the glass tube will break