Defender 2200 Bench Scale
EN-28
4.6.3 Display-Hold
Use this mode to capture and store the first stable display value.
To enter the Display hold mode, press and hold the
Mode
button until HOLd is
displayed.
When the
Mode
button is released, the Hold indicator will be turned on.
Place the item on the scale, The first stable value will be held on the display and the
Hold icon will blink. When a successive weight is added to the scale the display will
show the new value.
Even after removing the items,the weight will be held on the display.
To clear the held value, press
Mode
button.
NOTE
: Press and hold the
Mode
button to enter other modes.
4.6.4 Check Weighing
Use this mode to determine if the weight of a sample is within Pre-set limits.
Checkweighing Limits
When the
FUNCTION Mode
button is released, Clr.rEF is displayed.
Using or Recalling Stored Check Weighing Limits
Press the
No
button to recall the stored limits and proceed to checkweigh operation
Clearing Stored Check Weighing Limits
Press the
Yes
button to clear the stored limits.
NOTE
: Press the
FUNCTION Mode
button successively to temporarily display the
Under and Over Limit values.
to
Editing the Under Setting
The display shows SEt.LO. Press the
Yes
button to edit setting.
Settings:
–99999 to 999999
Refer to Menu Navigation Section 3.2 to enter settings.
NOTE
: The first digit will be used to show a negative value. Adjust the readability
setting as needed to allow an extra digit taken up by the negative sign
Содержание Defender 2200
Страница 1: ...Defender 2200 Bench Scale Instruction Manual...
Страница 2: ......
Страница 45: ......