Defender 2200 Bench Scale
EN-11
3.2 Menu
Navigation
TO ENTER THE MENU MODE
Press and hold the Menu button until MENU appears on the display. The first upper level menu
appears on the display.
Summary of button navigation functions in menu mode:
Yes
Allows entry into the displayed menu.
Accepts the displayed setting and advances to the next menu item.
No
Skips by the displayed menu.
Rejects the displayed setting or menu item and advances to the next
available
item.
Back
Moves backwards through the upper and middle level menus.
Backs out of a list of selectable items to the previous middle level menu.
Exit
Exits from menu directly to the active weighing mode.
For Checkweigh under and over targets setup, the current setting is displayed with all digits
flashing. Press the
No
button to begin editing.
The first digit is displayed flashing.
Press the
No
button to increment the digit or press the
Yes
button to accept the digit and move
to the next digit.
Repeat this process for all digits.
Press the
Yes
button when the last digit has been set.
The new setting is displayed with all digits flashing.
Press the
Yes
button to accept the setting or press the
No
button to resume editing.
3.3 Calibration
Menu
Span
Linearity
Geographic
Adjustment
End Calibration
Perform
Perform
0…31
End CALIBRATION menu
Two calibration processes are available: Span Calibration and Linearity
Calibration.
NOTES:
1. Make sure that appropriate calibration masses are available before
beginning calibration.
2. Make sure that the scale base is level and stable during the entire
calibration process.
3. Calibration is unavailable with LFT set to On.
4. Allow the Indicator to warm up for approximately 5 minutes after
stabilizing to room temperature.
5.
To abort calibration, press the Exit button anytime during the
calibration process.
Содержание Defender 2200
Страница 1: ...Defender 2200 Bench Scale Instruction Manual...
Страница 2: ......
Страница 45: ......