background image

Defender 2200 Bench Scale 

EN-33 

 

6. 

 

LEGAL FOR TRADE 

When the indicator is used in trade or a legally controlled application it must be set up, verified and sealed in 
accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that all 
pertinent legal requirements are met. 
 

6.1   Settings 

Before verification and sealing, perform the following steps: 

1. Verify that the menu settings meet the local weights and measures regulations. 
2. Perform a calibration. 
3. Set Legal for Trade to ON in the Setup menu. 
4. Exit the menu. 
5. Disconnect power from the indicator and open the housing as explained in Section 2.3.1. 
6. Set the position of the security switch to ON as shown in Section 1.2, Item 2. 
7. Close the housing. 
8. Reconnect power and turn the indicator on. 

 
NOTE: 

For installations that employ the audit trail sealing method, steps 5 to 8 are not required. However, the 

security switch may be set to ON to safeguard against unintentional changes to configuration and calibration 
settings. 
 

NOTE: 

When Legal for Trade is set to ON and the security switch is set to ON, the following menu settings cannot 

be changed: Span Calibration, Linearity Calibration, GEO, LFT, Calibration Unit, Capacity, Graduation, Power On 
Unit, Zero Range, Auto Zero Tracking, Expanded Mode, Count Mode, Kilogram Unit, Pound Unit, Gram Unit, 
Ounce Unit, Pound Ounce Unit, Stable Only. To enable editing of these menu settings, return the security switch to 
the off position and set LFT menu item to off. 
 

6.2   Verification 

The local weights and measures official or authorized service agent must perform the verification procedure. 
Please contact your local weights and measures office for further details.

 

Содержание Defender 2200

Страница 1: ...Defender 2200 Bench Scale Instruction Manual...

Страница 2: ......

Страница 3: ...libration 12 3 3 3 Geographical Adjustment Factor 13 3 3 4 End Calibration 13 3 4 Setup Menu 14 3 4 1 Reset 14 3 4 2 Legal for Trade 14 3 4 3 Calibration Unit 14 3 4 4 Capacity 14 3 4 5 Graduation 15...

Страница 4: ...Operation 25 4 3 Manual Tare 25 4 4 Changing Units of Measure 26 4 5 Printing Data 26 4 6 Application Modes 26 4 6 1 Weighing 26 4 6 2 Parts Counting 26 4 6 3 Display Hold 28 4 6 4 Check Weighing 28 4...

Страница 5: ...cal AC power Make sure that the power cord does not pose a potential obstacle or tripping hazard Use only approved accessories and peripherals Disconnect the equipment from the power supply before cle...

Страница 6: ...1 1 INDICATOR PARTS 1 Data Label 2 Front Housing 3 Adjusting Knob 2 4 Control Panel 5 Mounting Bracket 6 Screw 4 7 Rear Housing 8 Data Label 9 Power receptacle 10 Strain Relief for Load Cell Cable 11...

Страница 7: ...2 Overview of Parts and Controls Cont Figure 1 2 Main PC Board TABLE 1 2 MAIN PC BOARD Item Description 1 Line Power Input J1 2 LFT On Off Switch 3 RS232 Connector J3 4 Sense Jumper W2 5 Load Cell Ter...

Страница 8: ...1 Capacity Label Window 2 Pound indicator 3 Ounce 4 Kilogram gram indicator 5 TARE Menu button 6 CALIBRATION MODE indicator 7 FUNCTION Mode button 8 NET indicator 9 PCS indicator 10 PRINT Units button...

Страница 9: ...s Off Turns the Indicator off Units Changes the weighing unit Mode Allows changing the application mode Press and hold allows scrolling through modes Menu Enter the User menu Extended press View the A...

Страница 10: ...e power receptacle Figure 1 1 item 9 then plug the AC Adapter into an electrical outlet 2 2 3 Mounting Bracket to the Indicator Align the mounting bracket over the threaded holes in the side of the in...

Страница 11: ...d For a 6 wire load cell that includes sense wires see Figure 2 2 Jumpers W1 and W2 must be opened For load cells with an extra ground shield wire Connect the shield to the center position GND of J7 F...

Страница 12: ...READOUT AZT OFF 0 5d 1d 3d LINEARITY END END SPAN Y N Y UNIT MODE END END N N Y Y RESET YES NO CAPACITY 5 20000 GRADUATION 0 0005 20 CAL UNIT kg lb ZERO RANGE 2 100 RESET YES NO RESET YES NO KILOGRAM...

Страница 13: ...The first digit is displayed flashing Press the No button to increment the digit or press the Yes button to accept the digit and move to the next digit Repeat this process for all digits Press the Yes...

Страница 14: ...e exits to the active weighing mode and displays the actual weight value 3 3 2 Linearity Calibration Linearity calibration uses 3 calibration points The first calibration point is established with no...

Страница 15: ...tings from 0 to 31 are available Refer to table 3 2 to determine the GEO factor that corresponds to your location Note Changing the GEO Factor alters the calibration The GEO value was set at the facto...

Страница 16: ...17 18 18 19 19 20 20 21 21 22 22 23 23 24 24 25 25 26 26 27 27 28 28 29 29 30 30 04 04 05 05 06 06 07 07 08 08 09 09 10 10 11 11 12 12 13 13 14 14 15 15 16 16 17 17 18 18 19 19 20 20 21 21 22 22 23 23...

Страница 17: ...he Setup menu to the factory defaults No not reset Yes reset 3 4 2 Legal for Trade Set the legal for trade status OFF off ON on Turning on the LFT menu setting has the following effects Zero range is...

Страница 18: ...5 30 40 50 60 75 100 120 150 200 250 300 400 500 600 750 300 750 1000 0 05 0 1 0 2 0 5 1 0 2 0 5 1 5 10 15 20 25 30 40 50 60 75 100 120 150 200 250 300 400 500 600 750 1000 500 1000 1200 0 1 0 2 0 5 1...

Страница 19: ...f measure displayed at startup Auto last unit in use when the indicator was turned off Unit kg kilograms Unit lb pounds Unit g grams Unit oz ounces 3 4 7 Zero Range Set the percentage of scale capacit...

Страница 20: ...ision 2 d 2 divisions 3 d 3 divisions 3 5 3 Filter Set the amount of signal filtering LO less stability faster stabilization time MEd normal stability stabilization time HI greater stability slower st...

Страница 21: ...MEd Medium Intensity HI High Intensity 3 5 7 Auto Off Timer Set the automatic shut off functionality OFF disabled SEt 1 powers off after 1 minute of no activity SEt 2 powers off after 2 minutes of no...

Страница 22: ...Exit Mode menu 3 6 1 Reset Set the Mode menu to the factory defaults No not reset Yes reset 3 6 2 Parts Counting Mode Set the status OFF Disabled ON Enabled 3 6 3 Display Hold Mode Set the status OFF...

Страница 23: ...t Unit menu 3 7 1 Reset Set the Unit menu to the factory defaults No not reset Yes reset 3 7 2 Kilogram Unit Set the status OFF Disabled ON Enabled 3 7 3 Pound Unit Set the status OFF Disabled ON Enab...

Страница 24: ...o Interval 1 3600 Continuous Gross Off On Net Off On Tare Off On Unit Off On Exit PRINT menu 3 8 1 Reset Set the Print menu to the factory defaults No not reset Yes reset 3 8 2 Baud Set the Baud rate...

Страница 25: ...tAb printing occurs each time the stability criteria are met INtEr printing occurs at the defined interval CONt printing occurs continuosly When INtEr is selected set the Print Interval 1 to 3600 seco...

Страница 26: ...menu to the factory defaults No not reset Yes reset 3 9 2 Lock Calibration Set the status OFF Calibration menu is not locked ON Calibration menu is locked and hidden 3 9 3 Lock Setup Set the status O...

Страница 27: ...momentarily displays the software version and then enters the active weighing mode To turn the Indicator off press and hold the ON ZERO Off button until OFF is displayed 4 2 Zero Operation Zero can be...

Страница 28: ...he item to be weighed on the scale The illustration indicates a sample of 1 5 kg Gross weight Note To return to the Weighing mode from the other modes press and hold the Mode button until WEIGH is dis...

Страница 29: ...are 5 10 20 50 100 and 200 To establish the APW place the specified quantity of samples on the scale and press the Yes button to capture the weight Begin Counting Place the parts on the scale and rea...

Страница 30: ...4 6 4 Check Weighing Use this mode to determine if the weight of a sample is within Pre set limits Checkweighing Limits When the FUNCTION Mode button is released Clr rEF is displayed Using or Recalli...

Страница 31: ...to enter settings Begin Check Weighing The appropriate Under Accept or Over LED lights to indicate Check Weigh status Place a sample on the scale and read the weight For loads less than the Under Lim...

Страница 32: ...ess Mode button The Accum indicator will blink to show the value is added to the memory Remove the item scale must return to zero before the next item placed on the scale can be accumulated To display...

Страница 33: ...nt stable displayed weight according to stability setting CP Continuous Print SP Print when stable xP xP Interval Print x Print Interval 1 3600 sec Z Same as pressing Zero button T Same as pressing Ta...

Страница 34: ...Unit Stability Legend CR LF Length 1 1 7 1 5 1 3 1 1 Definitions Polarity sign if negative blank if positive Weight up to 6 numbers and 1 decimal right justified leading zero blanking Units up to 5 ch...

Страница 35: ...ing 8 Reconnect power and turn the indicator on NOTE For installations that employ the audit trail sealing method steps 5 to 8 are not required However the security switch may be set to ON to safeguar...

Страница 36: ...ictions that use the physical sealing method the local weights and measures official or authorized service agent must apply a security seal to prevent tampering with the settings Refer to the illustra...

Страница 37: ...Mode Kilogram Unit Pound Unit Gram Unit Ounce Unit Pound Ounce Unit Stable Only Note that the counter only indexes once even if several settings are changed The configuration event counter values rang...

Страница 38: ...esired weighing unit Unit not set to On Enable unit in the Units Menu Refer to Section 3 7 in the Unit Menu Cannot change menu settings Menu has been locked Lock switch set on Set selected menu to Off...

Страница 39: ...ct an authorized OHAUS Service Agent For Service assistance in the United States call toll free 1 800 526 0659 between 8 00 AM and 5 00 PM Eastern Standard Time An OHAUS Product Service Specialist wil...

Страница 40: ...71 mm 10 x 166 x 500 x 143 x 90 x 440 mm 10 x 166 x 500 x 143 x 90 x 440 mm Side View Dimensions S1 x S2 x S3 x S4 x S5 572 x 472 x 355 x 306 x 52 mm 920 x 668 x 550 x 500 x 52 mm 920 x 740 x 650 x 59...

Страница 41: ...pin 80500524 Interface Cable PC 9 pin 80500525 Column Kit 35 cm painted steel 80251743 Column Kit 70 cm painted steel 80251744 Column Kit 35 cm stainless steel 80251745 Column Kit 70 cm stainless ste...

Страница 42: ...t of Pan to Surface of Table Height of Pan Depth Including Level Indicator D30BRZ D60BRZ 355 mm 305 mm 123 mm 75 mm 380 mm D60BLZ D150BLZ 550 mm 420 mm 136 mm 80 mm 575 mm D150BXZ D300BXZ 650 mm 500 m...

Страница 43: ...rtified service organization established within the European Community or by the National Notified Body The first stage of the initial verification has been carried out at the manufacturer s work It c...

Страница 44: ...ent system is compliant with the ISO 9001 standard s requirements On May 15 2003 OHAUS Corporation USA was re registered to the ISO 9001 2000 standard LIMITED WARRANTY OHAUS products are warranted aga...

Страница 45: ......

Страница 46: ...oration 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 973 377 9000 Fax 973 944 7177 With offices worldwide www ohaus com 83033841 P N 83033841 2011 Ohaus Corporation all rights reserved Printed...