Remove any excess moisture from the cloth before wiping the cuff.
- Thoroughly wipe all surfaces of the cuff.
- Using a dry cloth, gently wipe off any excess moisture remaining on the cuff. Lay the
cuff flat in the unrolled position and allow the cuff to air dry.
Maintenance
- Never clean the device and the cuff with naphtha, thinner or petrol, etc.
- Do not wet the cuff or try to clean the cuff in water.
- Store the device in a clean, dry place.
- Do not subject the device to extremely hot or cold temperatures, humidity or direct
sunlight.
- Remove the batteries if the device will not be used for 1 month or more.
- The warranty is not valid if the problems encountered with the device are due to a
lack of maintenance or negligence not respecting the precautions for use and
maintenance indicated.
Exceptional situation
Error indicators
(Symbol appears on the screen during an abnormal measurement)
Symbol
Cause
Correction
E - 1
Weak signal or sudden
pressure change
Wrap the cuff correctly.
Measure again following the instructions.
E - 2
Strong external disturbance
Near a cell phone or other high radiation device, the
measurement will fail.
Stay silent and do not argue when taking the
measurement.
E - 3
An error appears during the
inflation process
Wrap the cuff correctly.
Make sure the air intake is properly inserted into the
device.
Measure again following the instructions.
E - 5
Abnormal blood pressure
Repeat the measurement after relaxing for 30 minutes. If
you get unusual readings 3 times, please contact your
doctor.
Low batteries
Replace all old batteries with new ones.
Elimination of problems
Problem
Check
Solution
Does not light
Replace the batteries
Replace the batteries
Installation and proper placement of
batteries
Installation and proper placement of
batteries
Does not swell
Insert firmly into the air intake
Insert firmly into the air intake
Change the cuff
Change the cuff
Err and goes
out
Rest the body
Rest the body
Keep calm while measuring
Keep calm while measuring
Leaked
armband
Tighten the cuff
Tighten the cuff
Change the cuff
Change the cuff
Please contact the distributor if you can not solve the problem, do not disassemble the unit