Maintenance
90
GA10218_002_A7 - 9/2015 - © Oerlikon Leybold Vacuum
4 Maintenance
4.1 Safety Information
Electrical hazards
During all maintenance and connection work, make sure that the mains cable have
been reliable disconnected and do not carry a mains voltage.The leak detector
must only be used in with the hoods closed. The electrical connections must only
be provided by a trained electrician as specified, for example, by the regulations EN
50110-1.
4.2 Maintenance Intervals
Maintenance work should be done on the PHOENIX L300i as required. This work
will normally be limited to exchanging the oil in the Trivac D2,5E rotary vane pump
and the built in air and oil filters.
As a preventive measure it is recommended that you check the rotary vane pump
once a month. Here note should be taken of the oil level and the colo ur of the oil.
Only Leybonol LVO 310 (Cat. no. L31001) must be used in the TRIVAC D2,5E in
the PHOENIX L300i.
The monthly interval for the check is just a nominal period. If the PHOENIX L300i is
used heavily, in particular in sniffer mode, then this check should be performed
more frequently. The rotary vane pump is located on the side of the mechanical
section at the bottom of the leak detector.
4.3 Oerlikon Leybold Vacuum Service
Whenever you send us in equipment, indicate whether the equipment is contami-
nated or is free of substances which could pose a health hazard. If it is contami-
nated, specify exactly which substances are involved. You must use the form we
have prepared for this purpose.
A copy of this form has been reproduced at the end of these Operating Instruc-
tions: Declaration of contamination for Compressors, Vacuum pumps and Compo-
nents. Another suitable form is available from www.oerlikon.com - Oerlikon
Leybold Vacuum Systems - Documentation - Download Documents.
Attach the form to the equipment or enclose it with the equipment.
This statement detailing the type of contamination is required to satisfy legal
requirements and for the protection of our employees.
We must return to the sender any equipment which is not accompanied by a con-
tamination statement.
Before shipping fit the yellow screw-on seals on to the connections EXHAUST and
GAS BALLAST.
Caution
Contamination
Form