![Oerlikon Leybold Vacuum PHOENIX L300i Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/oerlikon-leybold-vacuum/phoenix-l300i/phoenix-l300i_operating-manual_724987071.webp)
Operation
GA10218_002_A7 - 9/2015 - © Oerlikon Leybold
71
Partial flow setup/pump configuration for the PHOENIX L300i MODUL:
Before setting up the parameters the partial flow mode has to be enabled in the
main menu under Settings > Vacuum settings > Vacuum ranges with softkey no. 6
partial flow enable and confirm with the soft key OK.
Forepump setup for PHOENIX L300i MODUL
Softkey 2: Pump setup
Setting for the forepump if the PHOENIX L300i MODUL
is operated with a
partial flow system.
Softkey 7: Partial flow setup
Options for setting up the partial flow system.
Softkey 2:
The entry of the nominal pumping speed of the partial flow pump can be
decreased. The minimum pumping speed is 4 m
3
/h.
Softkey 3: Forepump type
The PHOENIX L300i MODUL with partial flow system can be operated
with a dry forepump (for ex. Scroll pump) or a wet forepump (oil sealed).
This key is for choosing a dry (Scroll, piston) fore vacuum pump.
Softkey 5: ?
Help text
Softkey 6:
The entry of the nominal pumping speed of the partial flow pump can be
increased. The maximum pumping speed is 80m
3
/h.
Softkey 7: Forepump type
The PHOENIX L300i MODUL with partial flow system can be operated
with a dry (for ex. Scroll pump) forepump or a wet pump (oil sealed).
This key is for choosing a wet (oil sealed) fore vacuum pump.
Partial flow setup PHOENIX
L300i MODUL
Vacuum settings PHOENIX
L300i DRY MODUL