Installation
32
GA10218_002_A7 - 9/2015 - © Oerlikon Leybold Vacuum
SERIAL/RS232
This RS232 interface Fig. 9/5 is wired as data communication equipment (DCE)
and permits the connection of a personal computer (PC) for monitoring and data
logging. The connection is made through a 9 pin sub-D socket. For more informa-
tion refer to chapter 3.5 and the Interface Description.
Remote Control (REMOTE1)
The remote control interface Fig. 9/10 is a serial interface to control the PHOENIX
L300i by the remote control. The remote control can be connected via an exten-
sion cable with a RJ45 plug. Refer to the Interface Description for more informa-
tion. The remote control does not belong to the standard equipment. If the remote
control is connected via a cable, wireless communication over REMOTE2 is
excluded.
REMOTE2
Through this interface the PHOENIX L300i is controlled wirelessly via Bluetooth or
WLAN. The Bluetooth transmitter connects to the remote control RC310 WL. The
WLAN module provides the connection to Apple I-Pad or other handheld devices.
Is the remote control RC310 WL connected via a cable to REMOTE1 REMOTE2
can not be used. The Bluetooth transmitter and the WLAN module are not included
PHOENIX L300i delivery.
Pin
Assignment
1
24 V switchable, default setting 0
2
TXD
3
RXD
4
GND 24V
5
GND
6
DSR
7
CTS
8
RTS
9
free
Pin
Assignment
2
+24V (fuse 0.8 A time lag)
3
0 V
4
RXD (intern. RS232)
5
TXD (intern. RS232)
SERIAL/RS 232
Remote control
REMOTE1
Remote control