Operation
GA10218_002_A7 - 9/2015 - © Oerlikon Leybold
67
The menu for the PHOENIX L300i MODUL version allows the following vacuum
settings which varies to the PHOENIX L300i:
Automatic purge
Main menu > Settings > Vacuum settings > Automatic purge
Through this menu is it possible to start automatic purge for 20 seconds automati-
cally.
This setting is only possible for the PHOENIX L300i MODUL (refer to vacuum set-
tings PHOENIX L300i MODUL)
Vent delay
Main menu > Settings > Vacuum settings > Vent delay
Through this menu item it is possible to define the delay time until the inlet port is
vented when operating the STOP button. When the STOP button is pressed for a
period of time which is shorter than the delay time specified here, the PHOENIX
L300i will just change to standby mode.
When the STOP button is pressed for a period of time which is longer than the
delay time specified here, the PHOENIX L300i will vent the inlet port.
Softkey 2: Automatic purge
Definition of automatic purge function in standby
Softkey 3: Vent delay
Definition of delay time until the PHOENIX L300i is vented
Softkey 4: Vacuum ranges
Selection of the vacuum working modes
Softkey 5: Partial flow setup/Forepump setup
Selection of partial flow mode and forepump type
Softkey 6: Sniffer factor
Setting for the sniffer factor
Softkey 7: Machine factor
Setting for the machine factor
Softkey 8: Leak rate internal test leak
Setting for the internal test leak
Softkey 3:
Off
The function automatic purge is off.
Softkey 6:
ON
The function automatic purge is on. When changing from measure to
standby the forevacuum pump will be purged automatically for 20 sec-
onds.
Vacuum settings PHOENIX
L300i DRY MODUL
Automatic purge
Vent delay