5-5. AUTO CURSOR MV
▌If the power supply doesn't receive any operation by using front-panel within a certain period of time,
the power supply's cursor location go down to minimum changing unit automatically
when you set the "AUTO CURSOR MV" mode is enable.
▌After passed long time, if a operator turns encoderder knob by mistake, cursor location of voltage and
current is at minimum unit position, so it can reduce the DUT damage.
▌Function
> DISABLE
Disable "AUTO CURSOR MV" mode.
> ENABLE
Enable "AUTO CURSOR MV" mode.
AUTO CURSOR MV Operating
▌Power switch on in a pressing "FACTORY" key.
▌Power switch ON
▌When appearing "1. user-mem clear" message, release "FACTORY" key.
▌Enter to "3. AUTO CURSOR MV" mode by using cursor key.
▌In order to set, press the "FACTORY" key one more time.
LCD Display
disable
▌Set the "enable" mode by using cursor key.
LCD Display
enable
▌To finish the "AUTO CURSOR" mode, press the "FACTORY" key one more time.
LCD Display
done
≫ Related Remote Interface Command
FACT:AUTO-CUR {DIS|ENA}
FACT:AUTO-CUR?
Return Value(0,1)
Apply: FACT:AUTO-CUR?
Return value : 1
"AUTO CURSOR' mode is enable state
FACT:AUTO-CUR DIS
"AUTO CURSOR' mode changed to