3-4. Programming Over Voltage Protection (OVP)
▌Turn on the power switch.
After turning on power, check if the unit is in "**OUTPUT OFF**" mode.
▌Press the protection key to set the levels.
Displayed LCD Message
PROTECT> 1.OVP
▌Press the protection key once more in "1.OVP" mode.
Displayed LCD Message
OVP SET> 32.00V
▌Use the cursor keys to place the cursor on the desired value.
▌Control the encoder switch to change the level value.
The OVP level cannot be set lower than the currently outputted voltage
to prevent operator mistakes.
▌Press the protection key when the change is complete.
Displayed LCD Message
CHANGED
Return to the previous stat after the message above is displayed.
Canceling
▌If you want to cancel the setting, wait for 5 seconds.
Displayed LCD Message
CANCEL…
≫ Related remote interface commands
VOLT:OVP {<voltage>}
VOLT:OVP?
VOLT:TRIP?
VOLT:CLE
Application: VOLT:OVP 25
Set OVP Level
VOLT:OVP?
View OVP Level
VOLT:OVP:TRIP?
Check OVP Trip Occurrence
VOLT:OVP:CLE
OVP Trip Clear
Refer to “Initial Setting Value” in “1-6 Check after Power On” for the initial OVP settings
This function protects the DUT from power supply if the output voltage is higher than the OVP
level. The procedure for setting OVP levels is described below.
You can disable the OVP
function in factory mode. The OVP level values shown below may be different from the actual
values.
Wait for 5
Seconds
Note