▌The LCD will display "
OVP TRIPPED
” if an OVP trip occurs.
▌To clear the OVP trip, start with increasing the OVP level.
Press the protection key to set the level.
Displayed LCD Message
protect> 1.ovp
▌Press the protection key once more in “
protect> 1.ovp
.”
Displayed LCD Message
ovp set> 15.00v
▌Use the cursor keys to place the cursor on the desired value.
▌Use the encoder switch to increase the level value.
▌Press the protection key when the change is complete.
Displayed LCD Message
CHANGED
After the message above is displayed, you will be returned to the state before
the trip occurred.
≫ Related remote interface commands
VOLT:OVP {<voltage>}
VOLT:OVP?
VOLT:OVP:TRIP?
VOLT:OVP:CLE
Application1:
Checking if a trip occurred and increasing the OVP level if it did
VOLT:OVP:TRIP? Return value "1"
Checking if there is an OVP trip
VOLT:OVP 32
Set OVP Level
VOLT:OVP:CLE
OVP Trip Clear
Application2:
Checking if a trip occurred and decreasing the OVP level if it did
VOLT:OVP:TRIP? Return value "1"
Checking if there is an OVP trip
VOLT 0
Setting the voltage to 0V
VOLT:OVP:CLE
OVP Trip Clear
- 26 -