When the power supply is connected from the input terminal of the electronic load,
voltage Regulation will occur in load connection lead wire.
To use more correct voltage, you can use Remote Voltage Sensing (V-Sensing).
Please refer following information when using V-Sensing
Connecting Remote Voltage Sensing
Set Remote Voltage Sensing mode. Connect sensing terminal to power supply terminal and
source output terminal to the load. Please be careful of +, - polarity.
< Figure 3-6 >
Remote Voltage Mode Setting
▌Power on, V_SENSING OFF
▌When INPUT is off(ON & SYN icons are off)
▌To enable V_SENSING ON, press SAVE/RECALL, V_SENSING key
for 2~3 seconds.
SYN icon is on.
▌To disable V_SENSING, press SAVE/RECALL, V_SENSING key
for 2~3 seconds.
SYN icon is off.
Note
V_SENSING setting can be saved when POWER is ON/OFF.
Even input comes to load's DC input terminal, voltage display will be displayed as 0
if there's no sensing wiring when V_SENSING is on.
- 58 -
3-14. Remote Voltage
POWER
SAVE/RECALL
V_SENSING
SAVE/RECALL
V_SENSING