BAT:RES?
Command for checking resistance value of BAT CR Mode.
Return value
"res"
ex) BAT:RES?
Return value "10.000"
BAT:WATT {VALUE}
Command for setting current value of BAT CP Mode. (Only available on BAT CP Mode)
>value
Input watt value
ex) BAT:WATT 10
SET watt value as 10W
BAT:WATT?
Command for checking resistance value of BAT CR Mode.
Return value
"watt"
ex) BAT:WATT?
Return value "10.000"
BAT:END:VOLT {VALUE}
Command for setting end voltage value of BAT Mode. (Only available on BAT Mode)
>value
Input voltage
ex) BAT:END:VOLT 10
SET 10V
BAT:END:VOLT?
Command for checking end voltage value of BAT Mode.
Return value
"voltage"
ex) BAT:END:VOLT?
Return value "10.000"
BAT:END:CAP {VALUE}
Command for setting end capacity value of BAT Mode. (Only available on BAT Mode)
>value
input capacity
ex) BAT:END:CAP 10
Set end capacity value as10mA
BAT:END:CAP?
Command for checking end capacity value of BAT Mode.
Return value
"capacity"
ex) BAT:END:CAP?
Return value "10.000"
BAT:END?
Command for checking "Pause State" by end condition of BAT Mode.
Return value
"0"
Before END (Working)
"1"
END (Not working)
- 91 -