ODA LF Series Скачать руководство пользователя страница 47

Cycling STEP DELAY SET

▌When in Cycling mode setting menu,
  CYC icon lights up and On icon lights off
▌Setting Cycling STEP DELAY menu
Move to Cycling STEP DELAY value menu by using left/right key
When in Cycling menu, "5.CYC_DELAY:" is displayed on 2nd line of  LCD
※Cycling setting munu order:  1.FINISH_STEP ↔ 2.REPEAT_NUM ↔ 
   3.MODE ↔ 4.VALUE STEP: ↔ 5.DELAY STEP: 

▌Set DELAY value by using encorder switch

▌Entering Cycling STEP DELAY menu 
 Press Cycling setting key once to enter Cycling STEP DELAY menu. 

▌To set Cycling STEP DELAY value, move cursor using right/left key
   to the value you want to change.
   And adjust the encorder switch to change value.

※10mS unit(Minimum 100mS) 
▌Saving Cycling STEP DELAY value
Press Cycling setting key.
After saving Cycling STEP DELAY value,
exit Cycling STEP DELAY and return to Cycling setting menu.

※ Repeat until each Step value is saved.

≫ Related Remote Interface Command
CYC:DELAY:{STEP_NUM} {VALUE} 
CYC:DELAY:{STEP_NUM}?

Exit Cycling mode setting menu

▌Exit Cycling mode setting menu
Press ESC & ERR, PROT key.
Exit Cycling mode setting menu and return to manual mode.

※ Exit after Cycling setting.

 -  38 - 

CYCLING 

SETTING

CYCLING 

SETTING

ESC

ERR/PROT

Содержание LF Series

Страница 1: ...www odacore com Programmable DC Electronic Load LF Series User Manual Manual Part NO 018LF 1 0 1...

Страница 2: ...d the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms in the separate agreement shall control Assist...

Страница 3: ...en Make all connections to the unit before WARNING applying power Note the instrument s A WARNING notice denotes a hazard It calls wire firmly connected to an electrical ground A CAUTION notice denote...

Страница 4: ...le push switch Out position of a bi stable push switch Caution risk of electric shock Earth ground terminal Protective earth ground terminal Frame or chassis terminal Terminal is at earth potential Ne...

Страница 5: ...heck 1 4 Operating Conditions 3 1 5 Check Before Power On 4 Power Cable AC Input Power 1 6 Check After Power On 4 Displayed proceduer on LCD Default Setting Values 1 7 Installation 5 Cooling Bench Ope...

Страница 6: ...nge LCD Display 3 5 ON OFF MODE 26 MODE SET Setting Value ON Current ON TIME OFF TIME ON OFF REPEAT Change LCD Display 3 6 FUSE MODE 29 MODE SET Setting Value FUSE Current FUSE TIME Change LCD Display...

Страница 7: ...MOTE Voltage 58 3 15 SHORT 59 4 CALIBRATION 4 1 Features 60 4 2 Preparation for Calibration 60 4 3 Equipment required for Calibration 61 4 4 Measurement Techniques 61 Instruments Connection Drawing Po...

Страница 8: ...74 5 5 CAL_BACKUP 75 5 6 CAL_DEFAULT 76 5 7 LOAD_DEFAULT 77 5 8 DELIMITER_CHAR 78 5 9 COM_RESPONSE 79 5 10 AOUT_KEY_LOCK 80 5 11 ADC_SAMPLING 81 6 1 Commands Syntax 82 6 2 Commands 82 7 1 Operating Er...

Страница 9: ...7 4 Calibration Error 102 7 5 Nonvolatile Memory Check Error 103 7 6 Interface Commands Error 104 8 Cautions 106...

Страница 10: ...nterface RS232C RS485 TCP IP Option SCPI Standard Commands for Programmable Instruments adopted High speed setting measument A large number of commands are programmed in product Easy interface setting...

Страница 11: ...5M Output cable 1 pc Rack mount User s Manual 1 pc Options RS232C Cable 1M 2M 4M RS485 Cable 2 4 8 Channel 1M 2M 4M 10M AC Input Cable Special order type Output Cable Special order type 1 2 Accessorie...

Страница 12: ...cal Check Check the broken key encoder switch power switch Check the panel surfaces are free of dents and scratches Check the broken rear output terminals Check the cabinet is free of scratches Check...

Страница 13: ...n LCD All icons on LCD and segments will light up ODA Technologies is displayed on line 1 The model name of LF Series is displayed on line 2 During displaying above procedure the instrument is initial...

Страница 14: ...ch Operation Load must be installed in a location that allows sufficient space at the sides front and rear of the unit for adequate air circulation There is no ventilation silt on the underside Rack M...

Страница 15: ...1 CP MODE LCD ICON 28 DYNAMIC RUN STOP Calibration KEY 12 CR MODE LCD ICON 29 DYNAMIC SETING Factory KEY 13 CC MODE LCD ICON 30 INPUT Terminal 14 CV MODE LCD ICON 31 INPUT Terminal 15 Setting Encoder...

Страница 16: ...N lights up when KEY LOCK mode In this mode you can t use any other keys 7 ERROR LCD ICON If errors occur the lamp lights up When all errors are checked the lamp goes out 8 V_SENSING ON LCD ICON light...

Страница 17: ...CC CV 22 IO LOCAL KEY As Remote Interface setting key you can select RS232C RS485 and TCP IP In Remote Interface mode you can change to Local Mode using this key 23 ESC ERROR PROTECTION KEY Can be us...

Страница 18: ...e During booting this key is for setting in Calibration menu and functional menu 29 DYNAMIC SETTING Factory KEY After booting the product this key is used to enter DYNAMIC MODE SETING menu During boot...

Страница 19: ...ngeable 50 60Hz is default and change the value is optional Before power on be sure to check AC switch on the bottom Default is AC220V 2 External Input Output Option Terminal 1 V_Sensing Input 2 V_Sen...

Страница 20: ...devices power supplier DMM Scope etc in parallel to one external controller Pin Description 1 5V Communication Power Source 3 D Data 2 D Data 4 0 Communication Ground 5 RS232C Interface Port This is a...

Страница 21: ...t the operation of constant power mode 3 5 ON OFF MODE Description about ON OFF Mode of constant current 3 6 FUSE MODE Description about FUSE Mode of constant current 3 7 BAT MODE BATTERY Description...

Страница 22: ...rors and disabling Protection 3 13 SAVE RECALL KEY LOCK Description about saving and loading the data from User Memory and disabling KEY LOCK mode 3 14 REMOTR Voltage Description about Remote Voltage...

Страница 23: ...AVE is displayed on the 2nd line of LCD present voltage and current will displayed on the 1st line and I SET will displayed on the 2nd line of LCD Related Remote Interface Command MODE CC CV CR CP ONO...

Страница 24: ...DISPLAY KEY and you can check various information when CC MODE is on DISPLAY change I SET WATT MODE TIME I SET When CC mode is on how to change current Check I SET setting When CC ON icon are on and...

Страница 25: ...NPUT OFF To disable CC MODE press INPUT ON OFF KEY If INPUT is off ON icon will light off Related Remote Interface Command INPUT OFF ON INPUT Note Day Hour Minute Second MD T Unit 000 00 00 00 16 INPU...

Страница 26: ...er SAVE is displayed on the 2nd line of LCD present voltage and current shows on the 1st line of LCD V SET will show on the 2nd line of LCD Related Remote Interface Command MODE CC CV CR CP ONOFF FUSE...

Страница 27: ...AY KEY and you can check various information when CV MODE is on DISPLAY change V SET WATT MODE TIME V SET When CV mode is on how to change voltage Check V SET setting When CV and ON icon are on and V...

Страница 28: ...NPUT OFF To disable CV MODE press INPUT ON OFF KEY If INPUT is off ON icon will light off Related Remote Interface Command INPUT OFF ON INPUT Note 19 Hour 000 00 00 00 Minute Second MD T unit Day INPU...

Страница 29: ...fter SAVE is displayed on the 2nd line of LCD present voltage and current shows on the 1st line R SET will show on the 2nd line of LCD Related Remote Interface Command MODE CC CV CR CP ONOFF FUSE BAT...

Страница 30: ...n DISPLAY change R SET WATT MODE TIME R SET When CR mode is on how to change resistance Check R SET setting When CR ON icon light on and R SET is displayed on the 2nd line LCD To set resistance move c...

Страница 31: ...PUT OFF To disable CR MODE press INPUT ON OFF KEY If INPUT is off ON icon will light off Related Remote Interface Command INPUT OFF ON INPUT Note Day Hour Minute Second 00 00 00 000 MD TIME Unit 22 IN...

Страница 32: ...u After SAVE is displayed on the 2nd line of LCD present voltage and current shows on the 1st line P SET will show on the 2nd line of LCD Related Remote Interface Command MODE CC CV CR CP ONOFF FUSE B...

Страница 33: ...P MODE is on DISPLAY change P SET MODE TIME P SET When CP mode is on how to change watt Check P SET setting When CP and ON icon are on and P SET is displayed on the 2nd line LCD To set resistance move...

Страница 34: ...PUT OFF To disable CP MODE press INPUT ON OFF KEY If INPUT is off ON icon will light off Related Remote Interface Command INPUT OFF ON INPUT Note MD TIME Unit 000 00 00 00 Second Hour 25 Day Minute IN...

Страница 35: ...UE menu MODE ON OFF is displayed on 1st line of LCD and ON VALUE is displayed on 2nd line of LCD To set ON_VALUE move cursor using right left key to the value you want to change And adjust the encorde...

Страница 36: ...d on 2nd line of LCD To set ON OFF_REP move cursor using right left key to the value you want to change And adjust the encorder switch to change value Saving MODE Exit MODE menu Press MODE key to save...

Страница 37: ...ON_VALUE WATT ON OFF TIME MODE TIME REPEAT ON_VALUE ON OFF MODE INPUT OFF To disable ON OFF MODE OFF press INPUT ON OFF KEY if INPUT is off ON icon will light off Related Remote Interface Command INPU...

Страница 38: ...MODE FUSE is displayed on 1st line of LCD VALUE is displayed on 2nd line of LCD To set FUSE value move cursor using right left key to the value you want to change And adjust the encorder switch to ch...

Страница 39: ...CC 1 CV 2 CR 3 CP 4 ON OFF 5 FUSE 6 BAT FUSE CURR VALUE FUSE CURR FUSE TIME VALUE FUSE TIME FUSE MODE INPUT ON To enable ON OFF MODE on press INPUT ON OFF KEY If INPUT is on ON icon will light up Rela...

Страница 40: ...is finished FUSE_MODE_END will be displayed on FUSE VALUE display and FUSE TIME will be stop and current value will be set as 0 When stop FUSE FUSE_CUT_OFF will be displayed on FUSE VALUE display and...

Страница 41: ...T MODE menu MODE BAT is displayed on 1st line of LCD and BAT_MODE is displayed on 2nd line of LCD To set BAT_MODE CC CR CCP move cursor using right left key to the value you want to change BAT_MODE ch...

Страница 42: ...LCD To set END CAPACITOR value move cursor using right left key to the value you want to change And adjust the encorder switch to change value Saving MODE Exit MODE menu Press MODE key to save MODE an...

Страница 43: ...WATT BT_CAP BAT_VALUE BATTERY MODE INPUT OFF To disable ON OFF MODE off press INPUT ON OFF KEY If INPUT is off ON icon will light off Related Remote Interface Command INPUT OFF ON INPUT All values in...

Страница 44: ...tep 1 FINISH_STEP is displayed on 2nd line of LCD Cycling setting munu order 1 FINISH_STEP 2 REPEAT_NUM 3 MODE 4 VALUE STEP 5 DELAY STEP Entering Finish Step menu Press Cycling setting key to enter Fi...

Страница 45: ...eturn to Cycling setting menu Related Remote Interface Command CYC REP VALUE CYC REP Cycling MODE SET When in Cycling mode setting menu CYC icon is on and On icon is off Cycling Mode menu setting Move...

Страница 46: ...e by using encorder switch Entering Cycling STEP value menu Press Cycling setting key once to enter Cycling STEP VALUE menu To set Cycling STEP value move cursor using right left key to the value you...

Страница 47: ...EP DELAY menu To set Cycling STEP DELAY value move cursor using right left key to the value you want to change And adjust the encorder switch to change value 10mS unit Minimum 100mS Saving Cycling STE...

Страница 48: ...STEP_VALUE Cycling mode STOP Check if Cycling is on CYC icon and On icon are light up To disable Cycling mode off press Cycling RUN STOP KEY CYC icon and On icon will light off and return to manual m...

Страница 49: ...right key When in Dynamic menu 1 MODE is displayed on 2nd line of LCD Dynamic setting menu order 1 MODE 2 A VALUE 3 B VALUE CYC TIME 5 A DUTY Entering Dynamic mode menu Press Dynamic setting key once...

Страница 50: ...r 1 MODE 2 A VALUE 3 B VALUE CYC TIME 5 A DUTY Dynamic A_VALUE menu Press Dynamic setting key once to enter Dynamic A_VALUE menu To set A VALUE move to REPEAT value which you want to set by using left...

Страница 51: ...UE CYC TIME 5 A DUTY Entering Dynamic B Value menu Press Dynamic setting key once to enter Dynamic B Value menu To set B VALUE move cursor using right left key to the value you want to change And adju...

Страница 52: ...UE CYC TIME 5 A DUTY Entering Dynamic CYC_TIME menu Press Dynamic setting key once to enter Dynamic CYC_TIME menu To set CYC TIME value move cursor using right left key to the value you want to change...

Страница 53: ...mic A DUTY menu To set A DUTY value move cursor using right left key to the value you want to change And adjust the encorder switch to change value B DUTY is 100 A DUTY Saving A DUTY value Press Cycli...

Страница 54: ...LAY change DYN A VALUE DYN B VALUE DYN CYC TIME A DUTY MD TIME DYN A VALUE Dynamic Mode STOP Check if Dynamic is on Dynamic icon and On icon are light up Dynamic mode STOP Press Dynamic RUN STOP KEY A...

Страница 55: ...4 CC L_CV L Saving CC CV RANG Press CC CV RANG KEY After saving CC CV RANG exit CC CV RANG menu and return to manual mode Note CC CV RANG KEY is only available when INPUT of CC CV mode is OFF Setting...

Страница 56: ...se check if the load is under KEY LOCK state RS 232C SET To set up RS 232C proceed as follows Power ON INPUT is off ON icon lights off Enter IO LOCAL menu Press IO Local Key to set RS 232C RS 232C men...

Страница 57: ...wiring drawing in case both cables are Female type of standard Cross cable Refer 1 2 Accessories and Options to select properly by length and use Figure 3 2 If the remote device equipped DB25PIN only...

Страница 58: ...nge RS 485 Baud Rate order BR_1 9600BPS BR_2 19200BPS BR_3 38400BPS BR_4 57600BPS BR_5 115200BPS Enter RS 485 ADDRESS menu Press IO Local key to set RS 485 ADDRESS Set RS 485 ADDRESS by using the enco...

Страница 59: ...ram of Remote device PC by ODA s 485 to 232 converter Figure 3 5 RS485 PC Interface Drawing RS485 of LF Series outputs DC5V 0 2A It doesn t need any extra power supply because this model is designed t...

Страница 60: ...der 1 RS 232C 2 RS 485 3 TCP IP Enter TCP IP menu Press IO Local Key to enter TCP IP menu Static IP setting Move to Static IP mode by using left right cursor When in Static IP mode TCP MODE_1 ST is di...

Страница 61: ...mode DHCP SET Power ON INPUT is off ON icon is off Enter IO LOCAL menu Press IO Local Key to set TCP IP TCP IP setting Move to TCP IP menu by using left right key When in TCP IP menu 3 TCP IP is displ...

Страница 62: ...nt Press IO Local Key to have TCP IP IP wait is displayed on the LCD Wait will be displayed while IP is being assigned After IP was assigned assigned IP will be displayed and Menu will be exited If IP...

Страница 63: ...ucture type From 11th error errors will be discarded in the order they occured After check the message pressing ERROR Key it will be deleted in the order errors occurred When ERROR occurs alarm and er...

Страница 64: ...on If heatsink s temperature is more than 100 Protection will be on and OVER TEMP PORT will be displayed To disable OTP stop operating load and cool down the load To disable OTP press ESC ERR PROT KEY...

Страница 65: ...displayed on 2nd line SAVE RECALL menu order 1 SAVE MODE 2 RECAL MODE Enter SAVE mode menu Press SAVE RECALL V_SENSING KEY to enter SAVE Mode Menu Set SAVE address number Select address number between...

Страница 66: ...MODE is displayed on 2nd line SAVE RECALL menu order 1 SAVE MODE 2 RECAL MODE Enter RECALL mode menu Press SAVE RECALL V_SENSING KEY to enter RECALL Mode Menu Setting RECALL address number Select addr...

Страница 67: ...e output terminal to the load Please be careful of polarity Figure 3 6 Remote Voltage Mode Setting Power on V_SENSING OFF When INPUT is off ON SYN icons are off To enable V_SENSING ON press SAVE RECAL...

Страница 68: ...ON icon lights up Press INPUT SHORT KEY LOCK KEY to enable SHORT MODE When SHORT ON SHORT ON is displayed on 2nd line Press INPUT SHORT KEY LOCK KEY to disable SHORT MODE off Then it return to CC MOD...

Страница 69: ...input 60 4 1 Characteristic 4 CALIBRATION you need to calibrate your electronic load according to the environment Connect each measuring instruments without any contact resistance Turn off the electro...

Страница 70: ...spec to make load s performace properly Measuring Instrument 4 3 Measuring instrument specifications Use Voltage Range 200 Vdc Current Range 60 Adc Based on LF 300A 61 Digital Voltmeter Voltage Calib...

Страница 71: ...rrent monitoring resistor Normal current measuring instrument is not suitable for high current capacity Select less than TCR 10ppm Use 0 01 high precise standard resistor DVM Digital Volt Meter Used f...

Страница 72: ...he front panel CALIBRATION KEY STRUCTURE 63 4 5 Calibration using front panel DYN RUN STOP CALIBRATE POWER ON 1 CC_HI_RNG DYN RUN STOP CALIBRATE 2 CC_LOW_RNG 3 CV_HI_RNG 4 CV_LOW_RNG CC_HI_LOW CC_HI_H...

Страница 73: ...ne Power on Release the key after model name is displayed and entered CALIBRATION mode Move to 1 CC_HI_RNG and press DYNRUN STOP Calibration Key then CC_HI LOW will be displayed CALIBRATION menu order...

Страница 74: ...ting press DYNRUN STOP Calibration key to save the setting value After HAX DAT is displayed go back to Calibration menu Move to 2 CC_LO_RNG and press DYNRUN STOP Calibration Key then CC_HI LOW will be...

Страница 75: ...lue is 0 001 and the measured voltage is 6 12mV the current value is 6 12A Input the calculated current value to the load using cursor key and encoder switch After setting press DYNRUN STOP Calibratio...

Страница 76: ...ey then CV_HI LOW will be displayed CALIBRATION menu order 1 CC_HI_RNG 2 CC_LOW_RNG 3 CV_HI_RNG 4 CV_LOW_RNG Wait until DVM s voltage becomes stable Input the calculated voltage value to the load usin...

Страница 77: ...RUN STOP Calibration key to save the setting value After HAX DAT is displayed it will be changed to CV_LOW_HI_CAL mode CV_LOW HI will be displayed Wait until DVM s voltage becomes stable Input the cal...

Страница 78: ...d to the load Command CAL CURR H Power supply s current value should be higher than load s max current value and set voltage as 5V If the power suppliy s current value is lower than load s calibration...

Страница 79: ...delay is more than 500mS After several minutes measure DVM voltage connected to Current Monitoring resistance Send measured value to the power supply s voltage value For example if measured value is...

Страница 80: ...et current several tens of mA If the voltage is less than or far greater than calibration range of the load CV can be measured incorrectly E g LF300 A CALIBRATION 155 160V 20 100mA After several minut...

Страница 81: ...ta and set to default value 5 FACTORY 72 5 2 FACTORY KEY Strecture 5 1 Features DYN SETTING FACTORY POWER ON 1 USER_CLEAR DYN SETTING FACTORY 2 CAL_RESTORE 3 CAL_BACKUP 4 CAL_DEFAULT 5 LOAD_DEFAULT 6...

Страница 82: ...and entered FACTORY mode Move to USER_CLEAR menu using cursor key 1 USER_CLEAR will be displayed FACTORY menu order 1 USER_CLEAR 2 CAL_RESTORE 3 CAL_BACKUP 4 CAL_DEFAULT 5 LOAD_DEFAULT 6 DELIMITER_CH...

Страница 83: ...ntered FACTORY mode Move to CAL_RESTORE menu 2 CAL_RESTORE will be displayed FACTORY menu order 1 USER_CLEAR 2 CAL_RESTORE 3 CAL_BACKUP 4 CAL_DEFAULT 5 LOAD_DEFAULT 6 DELIMITER_CHAR 7 COM_RESPONSE 8 A...

Страница 84: ...g DYN SETTING FACTORY key Power ON Release the key after model name is displayed and entered FACTORY mode Move to CAL_RESTORE menu 3 CAL_BACKUP will be displayed FACTORY menu order 1 USER_CLEAR 2 CAL_...

Страница 85: ...RY key Power ON Release the key after model name is displayed and entered FACTORY mode Move to CAL_DEFAULT menu using cursor key 4 CAL_DEFAULT will be displayed FACTORY menu order 1 USER_CLEAR 2 CAL_R...

Страница 86: ..._DEFAULT menu 5 LOAD_DEFAULT will be displayed FACTORY menu order 1 USER_CLEAR 2 CAL_RESTORE 3 CAL_BACKUP 4 CAL_DEFAULT 5 LOAD_DEFAULT 6 DELIMITER_CHAR 7 COM_RESPONSE 8 AUTO_KEY_LOCK 9 ADC_SAMPLING Pr...

Страница 87: ...STORE 3 CAL_BACKUP 4 CAL_DEFAULT 5 LOAD_DEFAULT 6 DELIMITER_CHAR 7 COM_RESPONSE 8 AUTO_KEY_LOCK 9 ADC_SAMPLING Press DYN SETTING FACTORY key Enter DELIMITER_CHAR menu D CHAR will be displayed Move to...

Страница 88: ...AL_DEFAULT 5 LOAD_DEFAULT 6 DELIMITER_CHAR 7 COM_RESPONSE 8 AUTO_KEY_LOCK 9 ADC_SAMPLING Press DYN SETTING FACTORY key Enter COM_RESPONSE menu RESP will be displayed Move to COM_RESPONSE using cursor...

Страница 89: ...6 DELIMITER_CHAR 7 COM_RESPONSE 8 AUTO_KEY_LOCK 9 ADC_SAMPLING Press DYN SETTING FACTORY key Enter COM_RESPONSE menu A LOCK will be displayed on 2nd line Set AUTO_KEY_LOCK using cursor key AUTO_KEY_LO...

Страница 90: ...LOAD_DEFAULT 6 DELIMITER_CHAR 7 COM_RESPONSE 8 AUTO_KEY_LOCK 9 ADC_SAMPLING Press DYN SETTING FACTORY key Enter COM_RESPONSE menu ADC_SAM will be displayed Set ADC_SAMPLING using cursor key ADC_SAMPL...

Страница 91: ...han one One command in one time Square bracket means option or parameters and it can be omitted Do not omit the parameter in bracket The value in Triangular bracket can be changed by CODE ex MIN MAX B...

Страница 92: ...ONOFF REP VALUE ONOFF REP ONOFF END FUSE CURR VALUE FUSE CURR FUSE TIME VALUE FUSE TIME FUSE END FUSE CUT BAT MODE CC CR CP BAT MODE BAT CURR VALUE BAT CURR BAT RES VALUE BAT RES BAT WATT VALUE BAT W...

Страница 93: ...MEM A DYN MEM B VALUE DYN MEM B DYN FREQ VALUE DYN FREQ DYN DUTY VALUE DYN DUTY Cycling Commands CYC INPUT ON OFF CYC INPUT ON CYC MODE CC CV CYC MODE CYC VAL STEP_NUM VALUE CYC VAL STEP_NUM CYC DELA...

Страница 94: ...Calibration Commands System Commands PROT PROT CLE SYST ERR IDN RST RCL VALUE SAV VALUE SN CAL VOLT MAX VALUE CAL CURR H L CAL CURR MIN VALUE CAL CURR MAX VALUE CAL VOLT H L CAL VOLT MIN VALUE 85...

Страница 95: ...nt Voltage MODE CR CR Mode Constant Resitstance MODE CP CP Mode Constant Power MODE ON OFF ON OFF Mode Constant Current ON OFF test MODE FUSE FUSE Mode Constant Voltage test MODE BAT BAT Mode Battery...

Страница 96: ...ltage ex VOLT Return value 10 000 RES value Command for setting input value of CR Mode Only available on CR Mode value Input resistance value ex RES 10 Set resistance as 10 ohms RES Command for checki...

Страница 97: ...ME as 1S Make a space after ONOFF ON TIME It is Hour Minute Second Millisecond from the beginning in order without a space ONOFF ON TIME Command for checking ON Time of ON OFF Mode Return value time e...

Страница 98: ...OFF Mode Return value 0 Before END Working 1 END Not working FUSE CURR VALUE Command for setting current value of FUSE Mode Only available on FUSE Mode value Input current value ex FUSE CURR 10 Set c...

Страница 99: ...y available on BAT Mode BAT MODE CC CC Mode BAT MODE CR CR Mode BAT MODE CP CP Mode BAT MODE Command for checking mode of BAT Mode Return value 0 CC Mode 1 CR Mode 2 CP Mode BAT CURR VALUE Command for...

Страница 100: ...value of BAT Mode Only available on BAT Mode value Input voltage ex BAT END VOLT 10 SET 10V BAT END VOLT Command for checking end voltage value of BAT Mode Return value voltage ex BAT END VOLT Return...

Страница 101: ...rn value 1 0000 MEAS WATT Command for measuring the output watt of load Return value watt ex MEAS WATT return value 1 0000 Dynamic Commands Command for Dynamic Mode DYN INPUT ON OFF Command to enable...

Страница 102: ...value 10 000 DYN MEM B VALUE Command for setting Bch HI value value Input Dynamic Bch value ex DYN MEM B 10 Set Dynamic as 10A or 10V Depending on selected Dynamic Mode CC CV voltage or current will...

Страница 103: ...INPUT ON OFF Command to enable or disable Cycling Mode ON Enable Cycling Mode OFF Disable Cycling Mode ex1 CYC INPUT ON ex2 CYC INPUT OFF CYC INPUT Command for checking if Cycling Mode is on or off Re...

Страница 104: ...rmation of Cycling Mode VALUE Cycling DELAY TIME ex CYC STEP DELAY 001 000000100 Save DELAY TIME 1S to Step 1 Set step 001 100 and make a space after CYC DELAY Input Hour 3digit Minute 2digit Second 2...

Страница 105: ...10 Set step between 1 10 CYC SEQ Command for checking last step of Cycling Mode Return value final step ex CYC SEQ return value 10 CYC FIN Command for checking if repeat time of Cycling Mode is finish...

Страница 106: ...ntering current High LOW Range Calibration set HI MAX current value VALUE ex CAL CURR MAX 60 123 CAL VOLT H L Command for entering voltage calibration H Command for HIGH RANGE voltage calibration L Co...

Страница 107: ...s can be applied to serial number for distrubution when develop Windows application Return value LF 00 0000 00000 ex SN return value LF 03 0923 00185 IDN Command for checking load s information Three...

Страница 108: ...latile User Memory 1 10 Memory save address ex sav 2 Save to address 2 RCL 1 2 3 4 5 6 7 8 10 Command for applying the data saved in non vilatile User Memory to your power supply Select one among 1 10...

Страница 109: ...is not operating 202 Front panel operating failed Front panel is not operating 255 Error not define Error occured but not defined Refer to 4 6 REMOTE INTERFACE Calibration for GPIB for further informa...

Страница 110: ...e Min VALUE MAX VALUE 28 Ignored min save under curr Occurs when operate MAX before operating Value of current MIN Procedure Min VALUE MAX VALUE 29 Invalid min value use under curr Occurs when operate...

Страница 111: ...low limit over Occurs when over low range of Voltage ADC high Rang 64 DAC V low Rang high limit over Occurs when over high range of Voltage DAC low Rang 65 DAC V low Rang low limit over Occurs when ov...

Страница 112: ...notify if an error occurred 80 Memory limit volt error There s an error in voltage limit value 81 Memory limit curr error There s an error in current limit value 82 Memory max volt error There s an e...

Страница 113: ...vised volt 10 122 Syntax error There s a syntax error ex volt Value should be added after volt Revised volt 10 123 Invalid suffix There s an error in suffix of received data ex volt 10 should be remov...

Страница 114: ...Revised volt 10 223 Incorret error Occurs when try a new operation without processing the buffer ex idn After send query command send a new command without acquiring data volt b data Revised idn a da...

Страница 115: ...not place liquid containers on top of product liquid can cause serious damage to the product Don t apply strong vibration or pressure to product Acquire enough space near the ventilation slit Don t pu...

Страница 116: ...62 Bupyeong daero 329beon gil Bupyeong gu Incheon city 403 858 Korea TEL 82 2 1800 8644 FAX 82 32 715 5456 www odacore com sales2 odacore com ODA TECHNOLOGIES CO LTD...

Отзывы: