11 Cleaning, Disinfection and Maintenance
38 / 55
Instruction Manual Binoptometer® 4P (G/59860/1016/en)
11 Cleaning, Disinfection and Maintenance
This chapter describes how to clean and disinfect the Binoptometer® 4P,
and how to replace the forehead rest and the fuses.
To ensure satisfactory and reliable operation, we recommend that you:
Have the Binoptometer® 4P checked every two years by our service
department or an authorized dealer.
If an error occurs that you can not resolve label the damaged
Binoptometer® 4P as "out of order” and report the damage to OCULUS
Service or your authorized dealer.
Use only undamaged Binoptometer® 4P devices.
Always heed the product descriptions and directions for use of
products you use to care for, clean, and disinfect the unit and/or its
accessories.
Do not clean the Binoptometer® 4P with aggressive, chlorinated,
abrasive or sharp cleaning agents.
11.1 Cleaning
Caution
Risk of electric shock if the Binoptometer® 4P is not completely
disconnected from the mains
Turn the Binoptometer® 4P off,
.
Pull the power plug before cleaning. When disconnecting electrical
connections, pull on the respective plug and not on the cable itself.
Materials needed:
Cleaner for plastic surfaces with anti-static effect
Cleaner for painted surfaces: Mixture of equal parts of alcohol and
distilled water, possibly with a few drops of commercial detergent
Soft cloth or lens brush
Alcohol or lens cleaner
Cleaning intervalls:
Wipe the forehead rest after each examination, and the housing as
required.