12 Transport and Storage
Instruction Manual Binoptometer® 4P (G/59860/1016/en)
43 / 55
12.2 Transport and Storage
Transport and storage conditions:
Wait approx. 3-4 hours after transport before operating the
Binoptometer® 4P for the first time. In the event of extreme
temperature changes from cold areas to warm rooms, the optical
components can become fogged.
Transport and storage conditions for daily transportation and short
storage times:
.
Note:
Equipment damage due to improper transport and storage
Avoid shocks, vibration and dust.
Avoid high temperatures and moisture.
Transport the Binoptometer® 4P in a compliant manner.
Store the Binoptometer® 4P in accordance with the storage
conditions.
Keep away from heating elements and moisture.
Check the Binoptometer® 4P for damage every time it has been
transported.
Ambient temperature range
-30°C to 70°C
Relative humidity range
10% to 90%,
Including condensation
Air pressure range
500 hPa to 1060 hPa
Ambient temperature range
-10°C to +40°C
Relative humidity range
30% to 75%,
Including condensation
Air pressure range
70 kPa to 106 kPa