Infrared
Probe
3
IR Channel -30ºF to 525ºF (-35ºC to 275ºC)
Temperature Range
35625-40 Food Safety Thermometer Specifications and Features
Accuracy
Assumes ambient operating
temperature of 73ºF ±4ºF
(23ºC ±2º)
Between 32ºF and 150ºF (0ºC to 65ºC): ±2ºF (±1ºC)
Below 32ºF (0ºC): ±2ºF (±1ºC) ±0.1 degree/degree
Above 150ºF (65ºC): ±1.5% of reading
<500ms after initial reading
Response Time
8-14 microns
Spectral response
Pre-set for Food Service Applications
Emissivity*
2.5:1 @ 90% energy, typical
Distance to Spot Size
Optical Resolution (D:S)
≈
0.5" to 10" (
≈
12mm to 250mm)
Typical Working Range
(target illumination)
0.5" (12mm)
Minimum Target Size
0.52" (13mm)
Illumination to IR
Channel Offset
-40ºF to 390ºF (-40ºC to 200ºC)
Temperature Range
Accuracy
Assumes ambient operating
temperature of 73ºF ±4ºF
(23ºC ±2º)
Between 23ºF and 150ºF (-5ºC to 65ºC):
±1ºF (±0.5ºC)
Below 23ºF (-5ºC): ±2ºF (±1ºC)
Above 150ºF (65ºC): ±1% of reading
7-8 seconds (3 time constants)
Response Time
Diameter: 0.118" (3.0mm) Length: 3" (78mm)
Probe Dimensions
Thin Film, Platinum Class A RTD
Sensor
4
Operational
Other
Within the accuracy specifications of the unit
Repeatability
35625-40 Food Safety Thermometer Specifications and Features
Ambient Operating Range
32ºF to 122ºF (0ºC to 50ºC)
90% (±5%) RH noncondensing @ 86ºF (30ºC)
Relative Humidity
-4ºF to 140ºF (-20ºC to 60ºC)
Storage Temperature
6.5" x 1.25" x 2" (165mm x 32mm x 50mm)
0.33lbs (150g)
Weight/Dimensions
(with Battery)
9V Alkaline
Power
10 hours minimum @ 73ºF (23ºC)
Battery Life (Alkaline)
High Brightness LED
Target Illumination
4 digits, 0.5ºF (0.2ºC)
Display Resolution
√
Display Hold (7 seconds)
Standards
Conforms to EN 61236-1 Electromagnetic
Emissions and Susceptibility, Criteria B, EN6101-
1 General Safety, IP54 Sealing (hand washable,
non-submersible)
CE
Certifications
1 year
Warranty
Nylon Carrying Pouch, QuickStart Guide
Accessories
√
LCD Backlit Display
*Calibration Geometry: >25C 140mm source @ 200mm (1:45:1) E=0.97 BB
<25C 140mm source @ 100mm (0.7:1) E=1.00 BB Specifications subject to change without notice