![NuAge Beauty COCOON AQUA IR HYDRATION PRO Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/nuage-beauty/cocoon-aqua-ir-hydration-pro/cocoon-aqua-ir-hydration-pro_owners-manual_1715629052.webp)
Section 11
System Cleaning, Maintenance and Reinstallation
Cocoon Aqua IR Hydration Pro System
48
11.4 Cleaning Contact Surfaces (Bed, Head Pillow, Interior of Shell)
1. Use Accel TB .05% Disinfectant on all contact surfaces AFTER EACH USE to present a
clean, sanitary appearance and feel for each new client.
2. Let solution sit for 60 seconds.
3. Wipe dry and clean with paper or soft cloth towel.
NOTE:
Cleaning after each use removes oils, perspiration, and cosmetics. It will also prolong
the material life of the accessories and bed.
11.5 Cleaning the Steam Generator
11.5.1: Steam Drainage Process
Recommended to be completed daily:
1. Turn the system On.
2. Press
Enter
to continue.
3. Use arrow keys to move to and highlight
Service
feature and press
Enter
.
4. Scroll down with arrow keys until
Drainage
feature is highlighted.
5. Press right arrow key to open drainage. You may hear a small click and water will begin
to flow from the nipple below the bed. Wait until water stops dripping.
6. Proceed by pressing left arrow key to adjust
Drainage
feature to “off.”
7. Press
Esc
. key to exit back to the main menu. Discard the excess water from drip pan.
This completes the drainage process. Leave unit empty of water each evening.
Refill water in AM before first use of the day.
11.5.2: Steam System Flushing Process
Recommended to be completed 2-3 times per week:
1. Turn the System On.
2. Press
Enter
to Continue.
3. Use arrow keys to move to and highlight
Service
feature and press Enter.
4. Scroll down with arrow keys until
Drainage
feature is highlighted.
5. Press right arrow key to open drainage. You may hear a small click and water will begin
to flow from the nipple below the bed. Wait until water stops dripping.
6. Proceed by pressing left arrow key to adjust drainage to off.
7. Press
Esc
key to exit back to the main menu. This completes the first round of system
flushing. Discard the excess water from the drip pan.
8. Unscrew the distilled water cap of the water reservoir.
9. Fill tank with room temperature distilled water until system indicates it is full with visual
message on LCD screen and audio beep.
10. Screw the cap back on to the distilled water reservoir.
11. Press
Enter
to continue.
12. Use arrow keys to highlight the
Program
feature and press
Enter
.
13. Attach a neck drape as if you were performing a full treatment. Roll a towel and place in
the area where head and neck of client would typically lie.