12
12
Model
Max. Rotation Speed (Motor)
Max. Rotation Speed (Handpiece)
Gear Ratio
Bur Type
Chuck Type
Chucking Length
Max. Bur Length
Max. Working Part Diameter
Use Environment
Transportation and Store
Environment
Thermo Disinfector
FX23
30,000 min
-1
30,000 min
-1
FX22
Latch Type
13.2mm
FX25
40,000 min
-1
40,000 min
-1
Push Button Type
12.7mm
FX25m
-
1:1 Direct Drive
ISO 1797-1 Type1 ø2.35mm CA Bur
22.5mm
ø4.0mm
Temperature: 10 - 40°C, Humidity: 30 - 75% (No Condensation)
Temperature: -10 - 50°C, Humidity: 10 - 85%,
Atmospheric Pressure: 500 - 1,060hPa
Points to check
Head cap or Cap (FX22) is loose
Rotation
Details
Tighten firmly using the correct head cap wrench or cap wrench.
Rotate the handpiece and check for abnormalities such as abnormal
rotation, vibration, noise, and overheating.
16. Specifications
Straight Handpiece (FX65/FX65m)
CAUTION
Ř
Read this Operation Manual before use to fully understand the product functions and file for future
reference.
Ř
When operating the product always consider the safety of the patient.
Ř
Users are responsible for the operational control, maintenance and continual inspection of this
product.
Ř
Do not disassemble or alter the handpiece except as recommended by NSK in this Operation Manual.
Ř
Do not allow any impact on to the product. Do not drop the product.
Ř
Operators and all others in the area must wear eye protection and a mask when operating this
handpiece.
Ř
Should the handpiece function abnormally during use, stop using the handpiece immediately and
contact your Authorized NSK Dealer.
Ř
Do not use the following fluids to wipe, immerse or clean the product; strong/super acid water,
strong acid/alkaline chemicals, chlorine-containing solutions, solvents such as benzine or thinner.
Ř
The handpiece is delivered in a non-sterile condition and must be autoclaved prior to use.
Ř
Perform regular function and maintenance checks. (Refer to "21. Periodical Maintenance Checks")
Ř
If the handpiece has not been used for a long period, rotate the handpiece and check for noise,
vibration and overheating before use.
Ř
To avoid clinical downtime it is recommended that a spare be kept on hand in case of a breakdown
during treatment.
Ř
U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
15. Periodical Maintenance Checks
Every 3 months perform periodical maintenance checks, referring to the check sheet below. If any
abnormalities are found, contact your Authorized NSK Dealer.