Specification
● 30
Nasal Cannula
Description
Properties
Physical
Filter
Hyrophilic Filer with Female Luer Lock Inlet
Environmental
Storage
Store in a cool dry place
Material
Cannula
PVC/Silicon
Filter
Acryl
Pulse Oximeter Module
Description
Properties
Range
Oxygen Saturation Range
(%SpO2)
0-100%
Pulse Rate Range
18-300 pulse per minute
Accuracy
Blood Oxygen Saturation
(%SpO2) ( ± 1 S.D.)
70-100% ± 2 digits for adults using 8000AA-WO Finger clip sensor
Pulse Rate
± 3%
Displays
Connection status LED
Amber or green
Recording time
Up to X hours
Measurement Wavelengths and Output Power
Red
660 nanometers @ mW nominal
Infrared
910 nanometers @ mW nominal
Internal Power
Battery
Two 1.5V AA batteries
Operating Life
Minimum 120 hours of continuous operation with new batteries
Storage Life
10 months
Weight
4.4 ounces with batteries (125g)
Environmental Condition
Temperature
Operation: 0°C to +50ºC (32°F to +122º F)
Storage/Transportation: -30°C to +50ºC (-22°F to +122º F)
Operating Altitude
Up to 40,000 feet
Humidity
Operation: 10-90% (relative humidity, non-condensing)
Storage/Transportation: 10-95% (relative humidity, non-condensing)
If transferred from a non-operating temperature and/or humidity condition,
allow at leas one hour of stabilization time before use.
Hyperbaric Pressure
Up to 4 atmospheres
Antenna Type
Inverted F type antenna
Antenna Gain
+ 2dB (typical), +3 dB (max)
Transmitter
Bluetooth®
compliance
Version 1.1
Operating frequency
2.4-2.4835 GHz
Output Power
< 1.1 mW
Operating Range
10 meter radius indoors
Network Topology
Point-to-Point
Operation
Slave: Model 4100
Antenna Type
Internal
Modulation Type
Frequency Shift Keying
Frequency Hopping Spread Spectrum
Bandwidth
1 MHz