Hook-Up
● 21
9.
To be able to use the pulse oximeter module in the recording it needs to be paired to the Nox
T3 device. Select the appropriate oximeter from the drop down list. In case the oximeter to
be used is not shown in the drop down list type in the ID code for that pulse oximeter,.
10.
Once the configuration is done press the
Configure Device
button. A
Device has been
configured
button will be displayed.
11.
Press the
OK
button and then disconnect the device.
Viewing Recorded Data
Before being able to view a recorded data in your PC you will need to upload the recording data to your
PC. Perform the following steps:
1.
Run the Noxturnal application
2.
Connect your device to a USB port on your PC using the USB cable delivered with the
system.
3.
The software automatically detects the device and uploads and analysis the recording. During
this task you can view various recording data in the
Review Study
sheet:
a.
Recording overview – an overview of the recorded signals and marked events for
each night recorded
b.
Subject information - shows the subject information entered when preparing the
study
c.
Recording statistics - a table of key sleep parameters that help with diagnosing the
recording data