![Nox Medical NOX T3 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/nox-medical/nox-t3/nox-t3_user-instruction_1713657023.webp)
Hook-Up
● 23
The
Print Report action
directly prints out the report above.
Hook-Up
It is assumed that the clinician will demonstrate the use of the device and applicable sensors to the
patient, or in the case of pediatrics to the parents, and the hook up then takes place at home by using a
detailed instruction for hook up. In some cases the clinician may prepare the patient partly or
completely.
Go through the following points with the patient:
1.
Applying of device and sensors
2.
Testing of sensor connections
3.
Status indications on the display including warnings
4.
Replacement of battery
It is important to remind the patient to follow the instructions given prior to the study.
Before sending the patient home:
1.
Make sure the device has been configured correctly
2.
Make sure the carry bag contains all equipment needed to finish the study, including batteries
3.
Provide the patient with the Nox T3 Hook Up Card
!
► Children should under no occasion hook up the system.
► Do not use damaged equipment like sensors or accessories.
► As with all medical equipment, carefully route cables and connections to reduce the
possibility of entanglement or strangulation
Attaching the Device and the Respiratory Effort Sensors
Step 1: Attach the device to the patient’s