background image

Pro-Meter V2K

 

 

 

 

 

 

 

6.  Be sure to tighten the stroke adjustment screw jam nut.

 

7.  Add material to the gravity tanks.

 

CAUTION:  

  Ensure the manual ball valves located just below the gravity tanks are CLOSED. When the ball 

valves are CLOSED, the handles are horizontal. 

  Fill each tank with at least a gallon of pourable material. 

 

If you have a gravity tank system, the dispense material should be pourable. If the material is 
particularly viscous, wait several hours to allow ingested air to emerge. 

 

 

 

 

 

1.  Install 6 to 8 inches (15 to 20 cm) of ¼ inch clear tubing to the bleed valve located on the high and low 

volume metering cartridge assembly. 

CAUTION: 

Utilize a cup, or suitable container, to collect bled material. 

NOTE:

 It is possible to add bled material back into the tank(s). 

 

 

 

2.  Utilizing the HMI pendant, proceed to the 

Level 1 Setup

 Screen and then to the 

Valve Setup 

Screen. 

Place the carriage in the METER FULL POSITION. 

  Select 

Open Inlet Valves 

pushbutton

.

 

  Select the Return Rods pushbutton to place the carriage in the METER FULL POSITION. 

3.  Position the ball valves located just below the gravity tanks to the OPEN position. 

NOTE: 

When the ball valves are OPEN, the handles are vertical. 

4.  Open bleed valves located on the metering cartridge assemblies. 

CAUTION: 

Bleed one side of the meter before proceeding to the next. 

NOTE:

 Open the bleed ports with a 7/16 inch wrench or a screwdriver. 

5.  Allow material to flow until air bubbles are no longer observed in the clear tubing. 

Bleed Procedure

 

NOTE:

 It may take longer to bleed viscous material. 

6.  Close bleed valves located on the metering cartridge assemblies. 

7.  Proceed to the 

Dispense 1 

Screen and note the following: 

CAUTION: 

Utilize a cup, or suitable container, to collect bled material. 

  Select 

Manual Mod

e pushbutton. 

  Dispense material via the dispense valve until air is no longer observed. 

  Proceed to 

Valve Setup 

Screen. 

  Place carriage in METER FULL POSITION. See Step 2. 

8.  Repeat steps 2 through 7 until air is fully bled from system. 

9.  Run a series of shots through material through the dispense valve to eliminate remaining air pockets. 

 

 

17

Содержание 7440272

Страница 1: ...application Dispense valve assembly Fitting kit TOOLS REQUIRED FOR INSTALLATION Ratcheting wrenches Adjustable wrenches Box cutter to cut crate tie straps Hammer to open crate Metric hex keys Metric s...

Страница 2: ...RNING Any material injected into flesh can cause severe injury or death Keep body parts awayfrom material discharge ports If injection occurs seek medical attention immediately WARNING Operating fault...

Страница 3: ...blue and P6 red female connectors located at the power supply Refer to the System Layout Drawing for additional information NOTE The HMI pendant harness connectors are labeled P5 and P6 and are unique...

Страница 4: ...ruction Sheet Nordson part number 7441862 for additional information 6 Ensure the gravity tank gaskets are compatible with the selected material 7 CAUTION Ensure the gravity tank gasket is seated even...

Страница 5: ...power switch located at the power supply to the ON position NOTE Refer to the Pro Meter V2K Screen Map in Appendix A for additional information 1 Proceed to Main Menu Screen 2 Select SETUP LOGIN pushb...

Страница 6: ...shbuttons Cylinder Setup 1 From the Main Menu Screen select the SETUP LOGIN pushbutton and proceed to the Setup Login Screen 2 From the Setup Login Screen select the LEVEL 1 SETUP pushbutton to procee...

Страница 7: ...ly close the air cylinder flow controls Slowly open the flow controls at turn intervals until the desired flow rate is achieved 3 From the Main Menu Screen proceed to the Setup Login Screen to the Lev...

Страница 8: ...he air cylinder is allowed to fully return as the RETURN RODS pushbutton is depressed 6 With the air cylinder fully extended verify the existence of a small gap between the stop block and the carriage...

Страница 9: ...READING data display into the FULL SCALE MM CAL VALVE data entry field NOTE The linear position sensor should now be calibrated 10 With the rods fully returned utilize a 6 mm wrench or equivalent too...

Страница 10: ...ure Sensor Calibration Screen 11 Enter the 0 psi Cal value and the Full Scale psi value 250 psi from the calibration sheet 12 Adjust the Zero Offset until the Scaled Pressure reads zero CAUTION If the...

Страница 11: ...cle will be run The purge timer will then reset Manual Shot Mode When the dispense command is given either by the foot switch or the DISPENSE pushbutton on the Dispense 1 Screen the Pro Meter V2K will...

Страница 12: ...our local Nordson Sealant Equipment Representative to order the purge cord Option 2 The customer purchases a jumper kit Nordson part number 7442948 and manually installs a jumper to pins 3 and 4 of th...

Страница 13: ...or fault The Pro Meter V2K will remain operational unless some other fault interferes with operation When the PURGE TIMER is turned ON the TIME LEFT TO PURGE will be displayed While in Manual Mode the...

Страница 14: ...supply line customer supplied 4 7440843 Purge ok splitter if supplied 5 7440844 Machine interlock cordset if supplied 6 7440845 Footswitch if supplied OR 7442945 Hand trigger if supplied 7 7442948 Pur...

Страница 15: ...close to 75cc as possible Stroke Setup 2 From the Level 1 Setup Screen proceed to the Valve Setup Screen 3 Depress the ADVANCE RODS pushbutton until the air cylinder is fully retracted Stroke Length S...

Страница 16: ...ch or equivalent tool and move the stop block until the gap is as close to 4 14 inch 10 52 cm NOTE It may be necessary to make further adjustments to fine tune the dispense volume It is also possible...

Страница 17: ...on Select the Return Rods pushbutton to place the carriage in the METER FULL POSITION 3 Position the ball valves located just below the gravity tanks to the OPEN position NOTE When the ball valves are...

Страница 18: ...nd Volume Adjustment Linear Speed Control Calibration 1 Utilize a 1 8 inch or 3 mm hex key adjust the air cylinder flow control valve in the open position WARNING Slowly adjust the flow control valve...

Страница 19: ...gently pry open low level sensor adjustment screw access cover B 2 Turn low level sensor adjustment screw A slowly counter clockwise with a small flat head screwdriver until the sensor turns ON 3 Tur...

Страница 20: ...APPENDIX A 20...

Страница 21: ...Issued 7 2018 Original copyright date 2017 Pro Meter and the Nordson logo are registered trademarks of Nordson Corporation www nordson com nse 1 800 433 9319 21...

Отзывы: