2 - 4
510
180
153
322
40
140
425
470
Opening A
Position of stage foot
(4 spots)
(Measure: mm)
30
50
Opening B
2.4 Installation
CAUTION
• Use the instrument under the following conditions.
- A dust free environment
- A stable place with no tilt
- A vibration and shock free environment
• Install the instrument where it can be used at the following temperature and humidity.
Use conditions
Temperature:
5 - 40ºC
Humidity:
Relative humidity does not exceed 50% at a
maximum temperature of 40ºC.
Pressure:
700 - 1060 hPa
• Install the instrument in a location that provides the operation area specified on p. 2-6.
• Install the instrument on the specified table (option) or a stage which is strong enough for the
instrument weight (42 kg).
Vibrations and noises may produce, or normal processings may not be performed.
• Use a stage whose size is larger than the
measurements of the figure to the right. Also
set the adjuster at the foot of the stage so that
the stage can be adjusted without tilt or play.
It is necessary to make openings in the top to
draw the pipe and cord through.
Opening A: For power cord of pump
Opening B: For feedwater pipe and drain pipe
• Do not install the instrument in a place exposed to direct sunlight or where the temperature and
humidity are high.
• Install the instrument where pollutants such as corrosive gases, acids, and salts are not present.
Corrosion or malfunction of the instrument may occur.
• Be sure to leave 3 cm or more of space in the rear so that the fan will not be blocked.
If the fan is blocked, the temperature of the inside of the instrument will rise and it may cause
malfunction.