86
DEVICE SETTINGS AND MAINTENANCE: Device Parameter Settings
●
Chart to be initially displayed (when “AUTO” is selected)
The measurement starting chart of uncorrected visual acuity measurement is calculated from AR values.
When cylindrical power is contained, -CYL is added to the spherical power.
When no AR data is contained, a full VA chart (VA 0.1 to 1.25) is displayed. AR measurements in the
table are values converted into 0.25 increments.
●
READER (Barcode scanner / Magnetic card reader function)
●
Checking the setting conditions in the ID field
When a barcode is read by the barcode scanner after setting, the read results under the set con-
ditions are displayed in the ID field. If anything other than alphanumeric characters, space, minus
sign, or underbar is entered, ~ (tilde) is displayed. Control codes are not displayed.
AR measured value
(D)
Chart to be presented
(Decimal display)
Chart to be presented
(Fraction display)
-0.25
≤
1.25
16
-0.50
1.0
20
-0.75 to -1.00
0.8
25
-1.25
0.63
30
-1.50
0.5
40
-1.75
0.4
50
-2.00
0.32
60
-2.25 to -3.00
0.25
80
-3.25
≥
< 0.1
> 200
99. ADD SELECT
(Addition power)
0.25D / 0.50D / 0.75D / 1.00D / 1.25D / 1.50D / 1.75D / 2.00D
Selects the addition power added by pressing the ADD
button in near vision
measurement mode.
100. GLARE CHART
(Glare chart)
CHANGE / STAY
Chart to be presented when the screen switches to the contrast/glare visual acuity
measurement.
CHANGE
: A VA chart one level lower is presented
STAY
: A VA chart at the same level is presented.
Parameter option
Setting contents
101. READER START
(Position to start reading)
1 to 250
Sets the position to start reading ID when the magnetic card reader is used.
Control codes are included in the number of characters. Set “1” when the barcode
scanner is used.
102. READER LENGTH
(Length of reading)
1 to 14
Sets the length of data to be read as ID when the magnetic card reader is used.
This reads the set length of data or reads up to the return code. Control codes are
not included in the number of characters. Set “14” when the barcode scanner is
used.
Change
Normal display
All data display
㹎㸿㹐㸿㹋㹃㹒㹃㹐ࠉࠉ㹑㹃㹒㹒㹇㹌㹅
ࠉࠉࠉࠉࠉ㹙㹐㹃㸿㹂㹃㹐㹛
㸯㸮㸯㹐㹃㸿㹂㹃㹐ࠉ㹑㹒㸿㹐㹒ࠉࠉࠉࠉࠉࠉࠉ㸯
㸯㸮㸰㹐㹃㸿㹂㹃㹐ࠉ㹊㹃㹌㹅㹒㹆ࠉࠉࠉࠉࠉ㸯㸲
ࠉࠉࠉࠉࠉ㹇㹂㸸
ࠉ㹆㹭㹱㹮㹇㹂ࠉ㹈㸿㹌㸮㸯㸯
㹼ࠉࠉࠉࠉࠉࠉ㹼
ᵏ
㹎㸿㹐㸿㹋㹃㹒㹃㹐ࠉࠉ㹑㹃㹒㹒㹇㹌㹅
ࠉࠉࠉࠉࠉ㹙㹐㹃㸿㹂㹃㹐㹛
㸯㸮㸯㹐㹃㸿㹂㹃㹐ࠉ㹑㹒㸿㹐㹒ࠉࠉࠉࠉࠉࠉࠉ㸯
㸯㸮㸰㹐㹃㸿㹂㹃㹐ࠉ㹊㹃㹌㹅㹒㹆ࠉࠉࠉࠉࠉ㸯㸲
ࠉࠉࠉࠉࠉ㸿㹊㹊㸸
ࠉ㸭㹆㹭㹱㹮㹇㹂㹜㹈㹎㹌㸮㸯㸯㸮㸰㸮㸯㸮㸮㸮㸰㸰㸭
㹎㹒㹇㹂㹜㸯㸭㹎㹒㹌㹓㹋㹜㸴㸮㸶㸲㸱ࠉࠉ㸭㹆㹭㹱㹮
㹇㹂㹜㹈㹎㹌㸮㸯㸯㸮㸰㸮㸯㸮㸮㸮㸰㸰㸭㹎㹒㹇㹂㹜㸯
㸭㹎㹒㹌㹓㹋㹜㸴㸮㸶㸲㸱
ࠉ㹜㸟㸨㸡㸢㸣̓㸸㸼㸺㹼㸪㸫㸬㸭㸮㸯㸰㸱㸲㸳㸴㸵㸶
㸷㸩㸹㸺㹜㸼㸽͇㸿㹀㹁㹂㹃㹄㹅㹆㹇㹈㹉㹊㹋㹌㹍㹎㹏
㹐㹑㹒㹓㹔㹕㹖㹗㹘͂㹹㸤㸻
ᵏ
Содержание ARK-1S
Страница 1: ...AUTO REF KERATOMETER AUTO REF KERATOMETER ARK 1s ARK 1s OPERATOR S MANUAL OPERATOR S MANUAL ...
Страница 6: ...IV ...
Страница 76: ...70 OPERATING PROCEDURE Operation when Peripheral Devices are Connected ...
Страница 112: ...106 SPECIFICATIONS AND TECHNICAL INFORMATION EMC Electromagnetic Compatibility ...