50
OPERATING PROCEDURE: Accommodation Measurement
4
Press the graph
button to display the graph
display screen.
The detailed graph, accommodation
*1
, maximum&minimum
AR-measured values, and maximum&minimum pupil size val-
ues are displayed.
The horizontal axis indicates the elapsed time in one-second
increments.
5
Press the graph
button to return to the accommodation measurement screen.
6
Switch the eye to be measured and measure the other eye in the same manner.
O
Graph display screen
*1. Accommodation increments is set by the “131. STEP” parameter independent from the display increments of the AR-measured
values.
ᵏ
$&&2002'$7,21
㸫㸷
㸷㸬㸯㸯
㸬㸯㸯
㸫㸯
㸯㸬㸴㸵
㸬㸴㸵
㸲㸬㸱
㸲㸬㸱
㸴㸬㸲
㸴㸬㸲
ᵘᵁᵚ
ᵘííᵚ
ᴴᴫᴱᴱ
• “ACC (accommodation)”, AR-measured “MIN (minimum value)”, “MAX (maximum value)”, pupil size
“MIN (minimum value)”, and “MAX (maximum value)” are printed out as measured results.
The greatest minus AR-measured value indicates “MAX (maximum value)”.
• To print the graph, set the “58. ACC GRAPH PRINT” parameter.
1
Accommodation (D)
2
Maximum AR-measured value (D)
3
Minimum AR-measured value (D)
4
Maximum pupil size (mm)
5
Minimum pupil size (mm)
6
Pupil size graph
7
Graph indicating chart position (TPOS)
8
Graph indicating change in AR-
measured values
(Accommodation graph)
•
Graph display of pupil size and AR-measured values
Interruptions on the graph indicate there were times that no measurement was obtained due to
blinking, alignment error, or such during measurement.
ᵏ
$&&2002'$7,21
㸫㸷㸬㸯㸯
㸯㸯
㸫㸯㸬㸴㸵
㸴㸵
㸲㸬㸱
㸴㸬㸲
ᵘᵁᵚ
ᵘííᵚ
ᴴᴫᴱᴱ
²